TECHNICAL DETAILS

  ActiveCount100 ActiveCount100H

Specifications

Flow Rate 100 liters/minute ± 4%
Impactor head 316L Stainless Steel 300 holes, 19 meters/second, autoclavable
Dust cover 316L Stainless Steel, autoclavable
Removable baseplate Magnetically attached removable, autoclavable
Adjustable baseplate Compatible with standard 90 mm petri dish (85mm-92mm), includes adjustment tool
Sample volume presets 8 programmable volume presets
User names 50 programmable user names (16 characters long)
Location names 400 programmable location names (16 characters long)
Display 3.5 inch (8.9 cm) color touch screen
LED light RGB LED light for status, alerts and alarms
Speaker Adjustable speaker volume
Time/Date formats Time format 12 or 24 hours, Date formats YYYY/MM/DD, MM/DD/YYYY, DD/MM/YYYY
USB port USB 2.0 port (8 GB flash drive inclued)
Data records Unlimited; saved on USB flash drive in .csv file format
Log file data fiels Start date, time, location, user, sample duration, flow duration, volume, cycles, status
Active flow control   Actively controled air flow at 100 liters per minute
Handle - tripod mountable   Tripod mountable handle 3/8-16, includes 1/4-20 threaded adapter
HEPA filterd exhaust   HEPA filtered exhaust captures 99.97% of particles as small as 0.3 µm
Enclosure 304 Stainless Steel 316L Stainless Steel
Handle - tripod mountable Tripod mountable handle 3/8-16, includes 1/4-20 threaded adapter
Power 24 VDC power adapter (100-240V ~ 1.5A 50-60Hz)
Battery 8 hours normal use, 4 hours charge time
Weight 5.80 lbs (2.63 kg)
Calibration Annual
Operating environment 0-90% humidity non condensing, 0-40 °C sampling, 0-35 °C charging
Dimensions 7.6”(h) x 4.7“(w) x 4.7”(d) [ 19.4 (h) x 12.0 (w) x 12.0 (d) cm]
Carry case Pelican case with Lighthouse logo

Included

  ActiveCount100, 304 stainless steel dust cover, 304 stainless steel sample head, petri dish adjust ment tool, 8 GB USB flash drive 1/4-20 tripod adapter, 24 volt power adapter, power chord, Opeating Manual on USB key, carry case  ActiveCount100H, 316L stainlees steel sample head, 316L stainless steel dust cover, removable baseplate, blower inlet seal, petri dish adjustment tool, USB 8 GB flash drive with cap, 1/4-20 threaded tripod adapter, 24 VDC power adapter with cord, operators manual on USB flash drive, calibration certificate, carry case.

Optional

  Stainless steel impactor head, cleanroom tripod, compressed gas sampler attachement, remote sample head, petri dish labels, IQ/OQ protocols Remote sampling adapter with iso probe, compressed gas sampler, tubing, cleanroom tripod, IQ/OQ protocols.

GET A QUOTE

First Name (*)
Last Name (*)
Company Name (*)
Email (*)
Phone (*)
Address (*)
Message (*)
Security Code
Security Code ImageRefresh Button
(*)
(*) Required Fields
ActiveCount100H

Microbial Sampler with Hepa Filtered Exhaust

ActiveCount100H is a high performance portable active air sampler suitable for use in cleanrooms and asceptic environments. The ActiveCount100H is truly ISO 14698 compliant with a d50 collection efficiency of 1 micron and HEPA filtered exhaust ports designed to avoid re-aspiration of sampled air. 

ActiveCount100H offers self-adjusting flow control to ensure accurate sampling as well as continuous & periodic sampling modes for optimum sampling flexibility. The autoclavable impactor head, dust cover and adjustable baseplate together with a sanatizable stainless steel enclosure and HEPA filtered exhaust; ensure ActiveCount100H will not contaminate your critical environments.

The optional remote sample head minimizes disturbance of unidirectional airflow within a BSC or laminar flow bench and the gas sample option provides the versatility needed to accomplish all your critical sampling requirements.

 

 

Key Features:

  • 3.5 inch (8.9 cm) color touch screen
  • ISO 14698-1 compliant
  • HEPA filtered exhaust
  • 8 programmable sampling volumes
  • 50 programmable user names
  • 400 programmable location names
  • Flow rate alarming
  • Data logging to USB flash drive
  • Password security for sample settings
  • Autoclavable sampling head and dust cover
  • Removable autoclavable petri dish holder (Patent Pending)
  • 24 VAC adapter or internal battery 8 hours normal use
  • Continuous, Periodic and Gas sampling modes
  • Wipedownable and suitable for sterile conditions