Remotes

Technical Details

  1104LD 2012 3012 5012 3102 5102 2014 3014 5014 3104 5104 5102V 5104V

Specifications

Size Range 0.10 - 1.0 μm 0.2 - 2μm 0.3 - 25μm 0.5 - 25μm 0.3 - 25μm 0.5 - 25μm 0.2 - 2μm 0.3 - 25μm 0.5 - 25μm 0.3 - 25μm 0.5 - 25μm
Channel Sizes Standard: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0 μm Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Standard: 0.5, 5.0μm
Standard: 0.5, 10.0μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Standard: 0.5, 5.0μm
Standard: 0.5, 10.0μm
Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Optional: 0.2, 0.3, 0.5, 1.0μm
Optional: 0.2, 0.3, 0.5, 2.0μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Optional: 0.3, 0.5, 1.0, 5.0μm
Standard: 0.5, 5.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm
Optional: 0.5, 5.0, 10.0, 25.0μm
Standard: 0.3, 0.5μm
Optional: 0.3, 0.5, 0.7, 1.0μm; 0.3, 0.5, 1.0, 5.0μm;
Optional: 0.3, 0.5, 5.0, 10.0μm
Standard: 0.5, 5.0μm; 0.5, 10.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm; 0.5, 5.0, 10.0, 25.0μm
Standard: 0.5, 5.0μm;
Optional: 0.5, 10.0; 0.5, 25.0 μm
Standard: 0.5, 5.0 µm
Optional 4 Channels: 0.5, 1.0, 5.0, 25.0 µm
Counting Efficiency 50% @ 0.1 μm; 100% for particles > 0.15 μm (per JIS) 50% @ 0.2μm; 100% for particles > 0.3μm (per JIS) 50% @ 0.3μm; 100% for particles > 0.45μm (per JIS) 50% @ 0.5μm; 100% for particles > 0.75μm (per JIS) 50% @ 0.3μm; 100% for particles > 0.45μm (per JIS) 50% @ 0.5μm; 100% for particles > 0.75μm (per JIS) 50% @ 0.2μm; 100% for particles > 0.3μm (per ISO 21501-4) 50% @ 0.3μm; 100% for particles > 0.45μm (per ISO 21501-4) 50% @ 0.5μm; 100% for particles > 0.75μm (per ISO 21501-4) 50% @ 0.3 μm; 100% for particles > 0.45 μm (per ISO 21501-4) 50% @ 0.5 μm; 100% for particles > 0.75 μm (per ISO 21501-4)
Flow Rate 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM)
Laser Source Long Life Laser Diode Laser Diode
Zero Count Level <1 count / 5 minutes (per JIS) < 1 count / 5 minutes (per ISO 21501-4)
Concentration Limits 500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss, per ISO 21501-4 2,000,000 Particles / ft3 @ 5% Coincidence Loss 500,000 Particles / ft3 @ 5% Coincidence Loss
Calibration NIST Traceable, per ISO 21501-4 NIST Traceable
Vacuum Requirements External Vacuum required, nominal 1.0 CFM >18" (45.72 cm) of Hg External Vacuum >18” (45.72 cm) of Hg
Air Flow - Internally Monitored Critical Orifice controlled Internally Monitored
Communication Modes RS 485 MODBUS 4-20mA RS-485 MODBUS 4-20mA RS-485/MODBUS
LED Indicators Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling
Data Storage - Rotating Buffer, 2000 Records - Rotating Buffer, 2000 Records
Supporting Software Lighthouse Monitoring System, LMS Express, LMS Express RT, LMS Express RT Plus Lighthouse Monitoring System LMS XChange, Lighthouse Monitoring System; LMS Express; Express RT; LMSNet
Environmental Sensors - (Optional) Temperature/Relative Humidity Probe 32 - 122F (0-50C)+/-1F (0.5C); 0 - 100% RH +/-2% -
External Alarm Output - Normally Open (NO), Dry-contact, Rated 0-60 VDC/VAC, 1 Amp -
Connector   Screw Terminal RJ45 Industrial RJ45
Enclosure Stainless steel 316L Stainless Steel, Wash Down Resistant 316L Stainless Steel, Wash Down Resistant/Splash-Tight
VHP Compatible - Yes
Power 24 VDC
Dimensions 10.5” (w) x 7.65” (h) x 3.35” (d) [26.67 x 19.43 x 8.51 cm] 1.7”(l) x 4.2”(w) x 2.5”(h) [4.3 x 10.7 x 6.3 cm] 3.5"(l) x 4.2"(w) x 3.3"(h) [8.89 x 10.66 x 8.38 cm] 1.7"(l) x 4.2"(w) x 3"(h) [4.3 x 10.7 x 7.6 cm] 5.4” (l) x 4.6” (h) x 1.65” (d) [13.72 x 11.68 x 4.19 cm] 4.2" (l) x 2.0" (w) x 3.7" (h) [10.7 x 5.08 x 9.398 cm]
Weight 7.0 lb (3.18 kg) 0.78 lb (0.35 kg) 0.97 lb (0.44 kg) 2 lb (0.90 kg) 0.97 lb (0.44 kg) 17.3 oz (0.49 kg) 15.6 oz (0.44 kg)

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Accessories

Included Operating Manual, Purge Filter, Power Supply, Power Cord Operating Manual on CD
Optional Isokinetic Sampling Probe, Sample Tubing, Cabling Printed Operating Manual; Isokinetic Sampling Probes; Vacuum Tubing; Sample Tubing; Data Cable; Purge Filter; Power Supply; High-Pressure Diffuser (not available for 2012); RS-232 Interface cable and adapter; USB to Serial Port Converter; REMOTE Mounting Bracket.  Printed Operating Manual; RS-485 Hub with Power Supply; RS-232 Communication Cable and Adapter; USB to Serial Port Converter; Network Adapter; Wireless Network Adapter; Isokinetic Sampling Probes; Vacuum Tubing; Sample Tubing; Data Cable; Purge Filter; Power Supply; Temperature/ Relative Humidity Probe; 8’ T/RH Probe Extension Cable; REMOTE Mounting Bracket; LMS XChange, LMS Express; Express RT, LMSNet software; RS-485 Communication Kit; High-Pressure Diffuser (not available for 2014) Isokinetic Sampling Probe, Sample Tubing, Cabling, Purge Filter, RS-485 Hub

Get a Quote

GET A QUOTE

First Name (*)
Last Name (*)
Company Name (*)
Email (*)
Phone (*)
Address (*)
Country (*)
Primary Industry (*)
Message (*)
(*) Required Fields

One for All

A remote for any application you can throw at us

Our line of remotes were built directly from customer feedback and those “special” applications that our customers challenge us with every day. We built this for you, with you, and because of our customers. Join the movement.

Instrument Files

Title
Download
Datasheet – Remote 2012, 2014, 3012, 3014, 3102, 3104, 5012, 5014, 5102, 5104
Download
Operating Manual – Remote 2012, 3012, 3102, 5012, 5102
Download