Remotes

Technical Details

  1104LD 2014 3014 5014 3104 5104 5104V

Specifications

Size Range 0.10 - 1.0 μm 0.2 - 2μm 0.3 - 25μm 0.5 - 25μm 0.3 - 25μm 0.5 - 25μm
Channel Sizes Standard: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0 μm Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Optional: 0.2, 0.3, 0.5, 1.0μm
Optional: 0.2, 0.3, 0.5, 2.0μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Optional: 0.3, 0.5, 1.0, 5.0μm
Standard: 0.5, 5.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm
Optional: 0.5, 5.0, 10.0, 25.0μm
Standard: 0.3, 0.5μm
Optional: 0.3, 0.5, 0.7, 1.0μm; 0.3, 0.5, 1.0, 5.0μm;
Optional: 0.3, 0.5, 5.0, 10.0μm
Standard: 0.5, 5.0μm; 0.5, 10.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm; 0.5, 5.0, 10.0, 25.0μm
Flow Rate 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM)
Laser Source Laser Diode
Zero Count Level < 1 Count / 5 minutes (meets specifications of JIS B 9921 and reporting requirements of ISO 21501-4 Annex C)
Concentration Limits 1,000,000 particles /ft3 (34,300,000/m3 ) @ 10% coincidence loss 4,000,000 Particles/ft3@10% Coincidence loss suitable for ISO classes 1-8 cleanrooms 1,000,000 Particles/ft3@10% Coincidence loss suitable for ISO classes 1-8 cleanrooms
Calibration NIST Traceable ISO 21501-4 Compliant and Accredited ISO 17025
Vacuum Requirements External Vacuum >18” (45.72 cm) of Hg
Air Flow Internally Monitored
Communication Modes Serial - 485 Modbus RJ45
LED Indicators Power, Flow, Service, Sampling
Data Storage Rotating Buffer, 2000 Records
Supporting Software LMS XChange, Lighthouse Monitoring System; LMS Express; Express RT; LMSNet
Environmental Sensors - (Optional) Temperature/Relative Humidity Probe 32 - 122F (0-50C)+/-1F (0.5C); 0 - 100% RH +/-2% -
Connector RJ45 Industrial RJ45
Power 24 VDC
Dimensions 10.5” (w) x 7.65” (h) x 3.35” (d) [26.67 x 19.43 x 8.51 cm] 1.7”(l) x 4.2”(w) x 3”(h) [4.3 x 10.7 x 7.6 cm] 1.7"(l) x 4.2"(w) x 3"(h) [4.3 x 10.7 x 7.6 cm] 1.7”(l)x4.2”(w)x3”(h) [4.3x10.7x7.6cm] 5.42” (l) x 2.14” (w) x 3.58” (h), [13.76 cm (l) x 5.43 cm (w) x 9.09 cm (h)]

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Get a Quote

One for All

A remote for any application you can throw at us

Our line of remotes were built directly from customer feedback and those “special” applications that our customers challenge us with every day. We built this for you, with you, and because of our customers. Join the movement.

Key Features:

  • 0.10 – 5.0 µm Sensitivity
  • 1.0 CFM (28.3 LPM) Flow Rate
  • Up to 8 Channels of Simultaneous Data
  • World’s Smallest 1.0 CFM 0.10 µm counter
  • Meets ISO 21501-4 Standards
  • Long Life Laser Diode
  • Small Footprint
  • Status Indicators
  • RS-485 MODBUS (R1100 Pulse output)
  • Stainless Steel Enclosure
  • Interfaces to Facility Monitoring Systems
  • Uses External Vacuum Source
  • Designed for Reliability
  • Designed to Ensure 21 CFR Part 11 Compliance when used with Additional Software
  • No Routine Laser Tube Replacements

Key Features:

  • 0.2 – 2.0 µm Particle Size Range
  • 0.1 CFM (2.83 LPM) Flow Rate
  • 4 Channels of Simultaneous Data
  • RS-485 Modbus Output
  • Meets 21501-4 Standards
  • Status Indicators
  • External Alarm Output
  • Stainless Steel Enclosure
  • Facility Monitoring Systems Interface
  • Versatile Mounting Options
  • External Vacuum Source
  • Designed for Reliability

Key Features:

  • 0.3 – 25 µm Particle Size Range
  • 0.1 CFM (2.83 LPM) Flow Rate
  • 4 Channels of Simultaneous Data
  • RS-485 Modbus Output
  • Meets 21501-4 Standards
  • Status Indicators
  • External Alarm Output
  • Stainless Steel Enclosure
  • Facility Monitoring Systems Interface
  • Versatile Mounting Options
  • External Vacuum Source
  • Designed for Reliability

Key Features:

  • 0.3 – 25 µm Particle Size Range
  • 1.0 CFM (28.3 LPM) Flow Rate
  • 4 Channels of Simultaneous Data
  • RS-485 Modbus Output
  • Status Indicators
  • Meets ISO 21501-4 Standards
  • External Alarm Output
  • Facility Monitoring Systems Interface
  • Stainless Steel Enclosure
  • Versatile Mounting Options
  • External Vacuum Source
  • Designed for Reliability

Key Features:

  • 0.5 – 25 µm Particle Size Range
  • 0.1 CFM (2.83 LPM) Flow Rate
  • 4 Channels of Simultaneous Data
  • RS-485 Modbus Output
  • Meets 21501-4 Standards
  • Status Indicators
  • External Alarm Output
  • Stainless Steel Enclosure
  • Facility Monitoring Systems Interface
  • Versatile Mounting Options
  • External Vacuum Source
  • Designed for Reliability

Key Features:

  • 0.5 – 25 µm Particle Size Range
  • 1.0 CFM (28.3 LPM) Flow Rate
  • 4 Channels of Simultaneous Data
  • RS-485 Modbus Output
  • Meets 21501-4 Standards
  • Status Indicators
  • External Alarm Output
  • Stainless Steel Enclosure
  • Facility Monitoring Systems Interface
  • Versatile Mounting Options
  • External Vacuum Source
  • Designed for Reliability

Key Features:

  • VHP Compatible
  • Class 1 Division 2 Rating
  • Built-in Flow Monitoring
  • 0.5 – 5.0 µm
  • 1.0 CFM (28.3 LPM) Flow Rate
  • Up to 4 channels of simultaneous data
  • Meets ISO 21501-4 standards
  • Real-time continuous monitoring
  • Small size
  • 316L Stainless Steel, Wash Down Resistant / Splash-Tight Enclosure
  • Status Indicators
  • RS-485 Modbus
  • Interfaces to Facility Monitoring Systems
  • Uses External Vacuum Source
  • Designed for Reliability
  • 2 Year Warranty

Instrument Files

Title
Download
Operating Manual – Remote 1100, 1104 LD
Download
Datasheet – Remote 1100,1104LD
Download
Operating Manual – Remote 5104V
Download
Datasheet – Remote 5104V