TECHNISCHE DETAILS
1100 LD | 3100 | 5100 | 3350 | Boulder Counter | |
Specifications |
|||||
Size Range | 0.1 - 1.0 μm Optional ranges up to 5.0 µm |
0.3 - 25.0μm | 0.5 - 25.0μm | 0.3 - 25.0 µm | 5.0 - 100.0 um |
Channel Sizes | Standard 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0μm Optional 8 channels: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 1.0, 5.0 µm |
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm Standard: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm 8-Channel: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0,7.0,10.0 µm |
Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm Standard: 0.5, 1.0, 3.0, 5.0, 10.0, 25.0 µm 8-Channel: 0.5, 0.7, 1.0, 2.0, 3.0, 5.0, 7.0,10.0 µm |
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm Optional: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm |
Standard: 5.0, 10.0, 25.0, 40.0, 50.0,100.0µm |
Flow Rate | 1.0 CFM (28.3 LPM) | 100 LPM | 28.3 LPM | ||
Counting Efficiency | 50% @ 0.1 µm; 100% for particles > 0.15 µm per ISO 21501-4 | 50% @ 0.3 µm; 100% @ sizes > 0.45 µm |
50% @ 0.5 um; 100% @ sizes > 0.75 µm |
50% @ 0.3 µm; 100% @ sizes > 0.45 µm | 50% @ 5.0 um |
Concentration Limits | 500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss | 1,000,000 particles/ft3 (35,300,000/m3) @ 10% coincidence loss Suitable for ISO classes 1-8 cleanrooms |
400,000 particles/ft3 (14,120,000/m3) @ 10% coincidence loss Suitable for ISO classes 1-8 cleanrooms |
500,000 particles/ft3 @ 5% coincidence loss | |
Light Source | Long Life Laser Diode | Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF) | |||
Dimensions | 14.65"Lx8.0"Wx10.4"H (37x20x26cm) | 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) | 14.65"Lx8.0"Wx10.4"H (37x20x26cm) | 9.5"Lx8.0"Wx10.2"H (24.13x20.32x25.9cm) | |
Weight | With battery: 23.55 lbs (10.68 kg) Without battery: 20.45 lbs (9.28 kg) |
With battery: 13.40 lbs (6.08 kg) Without battery: 12.00 lbs (5.44 kg) |
With battery: 26.15 lbs (11.86 kg) Without battery: 23.05 lbs (10.46 kg) |
With battery: 13.40 lbs (6.08 kg) Without battery: 12.00 lbs (5.44 kg) |
|
Zero Count Level | < 1 count / 5 minutes (per JIS B 9921) | ||||
Calibration | Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter | ||||
Count Modes | Automatic, Manual, Beep, Concentration, Cumulative/Differential, Real Time (with LMSNet or LMS Express RT software) | ||||
Data Storage | 3,000 Sample Records, Rotating Buffer (particle and environmental data with location and time) | ||||
Communication Modes | USB, Ethernet, RS-485 and Secure Data Transfer (LSDT) to and Storage on USB Flash Drive | ||||
Supporting Software | LMS Express and Express RT; LMS Pharma | ||||
Environmental Sensors | Optional (up to 4) temperature/relative humidity, air velocity, differential pressure or others | ||||
Touch Screen Display | 5.7-inch (14.47 cm) Color TFT Touch Screen Display | ||||
Reports | Pass/Fail Report Wizard for ISO-14644-1, EU GMP Annex 1 and FS-209 Report Formats | ||||
Alarms | Integrated alarm buzzer and external alarm connector. Alarms on counts, low battery, sensor failure | ||||
Sample Output | Internally Filtered to 100% @ 0.1μm, Optional HEPA Filtration for Cooling System Exhaust | Internally filtered to HEPA standards (>99.97% @ 0.3μm) | |||
Vacuum Source | Internal closed-loop controlled pump. Allows the use of 3/8" ID sample tubing of up to 12 meters (40 feet) @ 50 & 100LPM | ||||
Enclosure | Stainless Steel | ||||
Power | 100-240 VAC, 50-60 Hz, External AC/DC Adapter | ||||
Battery | Optional: Li-Ion, Removable & Rechargeable (typical recharge time 7.5 hours) | ||||
Languages | English (U.S.), French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean | ||||
Environmental Conditions |
|||||
Operating | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing | ||||
Storage | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing | ||||
Accessories |
|||||
Included | Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Power Cable and Spare Fuse; LMS XChange and Express Software; Printer; Printer Paper | ||||
Optional | Printed Operating Manual; Network Adapter; Wireless Network Adapter; Internal Removeable, Rechargeable Battery; Carrying Case with Wheels; Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe; High Pressure Difuser; 32-Port Universal Manifold; LMS Express RT Software; LMSNet Software; Validation Documentation | Wireless Network Adapter; Removable Rechargeable Li-Ion Battery; Carrying Case with Wheels; Temperature/Relative Humidity Probe; Differential Pressure Probe; Intelligent Filter Scanning Probe; Air Velocity Probe; Spare Removable Rechargeable Li-Ion Battery; External Battery Charger; LMS Express, RT Software; LMS Pharma Edition Software; LMS Standard Software; Validation Documentation; Printed Operating Manual; Cleanroom Cart |
OFFERTE AANVRAAG
Ongekende veelzijdigheid
Uw Solair Particle Counter biedt veel mogelijkheden
De Solair is de beste draagbare deeltjesteller op de markt, en kan verbonden worden met een manifold en gebruikt worden in een realtime monitoring systeem. Met de aanvulling van een ScanAirPro Set kunnen filter integriteitstesten uitgevoerd worden. Een optionele high pressure controller maakt metingen in (ultraschone) gassen mogelijk. Hiermee beschikt u over ’s werelds meest veelzijdige draagbare deeltjesteller

PRODUCT CONFIGURATOR
Filter Models and Find Your Model Number
Minimum Channel Size 2nd Selection
Flow Rate

Solair 1100LD
The SOLAIR 1100LD (Laser Diode) leads the industry as the first of its kind portable particle counter with 0.1 micron sensitivity and...

Solair 3100
Number 1 in Versatility The SOLAIR 3100 is the most versatile particle counter on the market today. Whether...

Solair 3350
The SOLAIR 3350 incorporates our extreme life laser diode technology sensor to produce the industry’s longest laser diode life of 20+...