Remote 5102PN
TECHNISCHE DETAILS
2014P | 3014P | 5014P | 3104P | 5104P | 5012PN | 5014PN | 5102PN | 5104PN | |
Specifications |
|||||||||
Size Range | 0.2 - 2.0 μm | 0.3 - 25.0 μm | 0.5 - 25.0 μm | 0.3 - 25.0 μm | 0.5 - 25.0 μm | ||||
Channel Sizes | Standard: 0.2, 0.3, 0.5, 1.0 μm | Standard: 0.3, 0.5, 1.0, 5.0 μm | Standard: 0.5, 1.0, 5.0, 10.0 μm | Standard: 0.3, 0.5, 1.0, 5.0 μm | Standard: 0.5, 1.0, 5.0, 10.0 μm | Standard: 0.5, 5.0μm Optional: 0.5, 10.0μm |
|||
Counting Efficiency | 50% @ 0.2 μm; 100% for particles > 0.3 μm (per ISO 21501-4) |
50% @ 0.3 μm; 100% for particles > 0.45 μm (per ISO 21501-4) |
50% @ 0.5μm; 100% for particles > 0.75μm (per ISO 21501-4) | 50% @ 0.3 μm; 100% for particles > 0.45 μm (per JIS) |
50% @ 0.5 μm; 100% for particles > 0.75 μm (per JIS) | 50% @ 0.5μm, 100% for particles >0.75μm (per JIS) | 50% @ 0.5μm, 100% for particles >0.75μm (per ISO 21501-4) | 50% @ 0.5μm, 100% for particles >0.75μm (per JIS) | 50% @ 0.5μm, 100% for particles >0.75μm (per ISO 21501-4) |
Flow Rate | 0.1 CFM (2.83 LPM) | 1.0 CFM (28.3 LPM) | 0.1 CFM (2.83 LPM) | 1.0 CFM (28.3 LPM) | |||||
Laser Source | Laser Diode | ||||||||
Zero Count Level | <1 count / 5 minutes (per ISO 21501-4) | <1 count / 5 minutes (per JIS); Zero Count with Vibration: 0.02g2/Hz from 8-66 Hz | <1 count / 5 minutes (per JIS) | ||||||
Concentration Limits | 2,000,000 Particles / ft3 @ 5% Coincidence Loss | 500,000 Particles / ft3 @ 5% Coincidence Loss | 2,000,000 particles / ft³ @ 5% coincidence loss | 500,000 particles / ft³ @ 5% coincidence loss | |||||
Calibration | Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter | NIST Traceable | |||||||
Sample Output | - | Internally filtered to HEPA standards (>99.997% @ 0.3μm) | |||||||
Vacuum Requirements | Internal Vacuum Pump | ||||||||
Air Flow | Internally Monitored and Controlled | ||||||||
Communication Modes | RS-485 MODBUS, Ethernet, USB | Ethernet, RS-485 MODBUS, Serial USB | 4-20mA Output | MODBUS: RS-485; Optional Ethernet MODBUS TCP | 4-20mA Output | MODBUS: RS-485; Optional Ethernet MODBUS TCP | |||
LED Indicators | Power, Flow, Service, Sampling | Power, Service | Power, Flow, Service, Sampling | Power, Service | Power, Flow, Service, Sampling | ||||
Data Storage | 3,000 Samples | - | Rotating Buffer, 2.000 Records | - | Rotating Buffer, 2.000 Records | ||||
Connector | RJ45, USB (Type B) | 2 RJ45, 2 RJ12 and one USB ‘B’ | 8-pin Phoenix | RJ45 | |||||
External Pump Switch | - | 2-pin Phoenix dry-contact relay | - | - | |||||
Supporting Software | LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and LMSNet | LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and RT+; LMSNet | Lighthouse Monitoring System; LMS Xchange | Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT | Lighthouse Monitoring System; LMS Xchange | Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT | |||
Up to Four 4-20 mA Inputs | Can be used with the Lighthouse standard 4-20 mA Temperature/Relative Humidity, Air Velocity or Differential Pressure Probes or any other 4-20 mA device | - | |||||||
External Alarm Output | - | Normally Open Dry Contact Rated 0-60 V AC/DC 1 Amp | - | ||||||
Enclosure | Stainless Steel | NEMA Style 316L Stainless Steel or Powder Coated Steel | |||||||
Power | 24 VDC | 110 / 220 VAC, 50-60Hz or 24VDC 100W | |||||||
Dimensions | 7.25”(l) x 3.75”(w) x 6”(h) [18.41 x 9.52 x 15.24 cm] | 9.5”(l) x 6.5”(w) x 8.51”(h) [24.13 x 16.51 x 21.62 cm] | 10” (w) x 12” (h) x 6.9” (d) [25.4 x 30.4 x 14.9 cm] | ||||||
Weight | 5.0 lbs (2.26 kg) | 12.33 lbs (5.59 kg) | AC Model, 20 lbs (9 kg); DC Model, 19.5 lbs (8.85 kg) | ||||||
Environmental Conditions |
|||||||||
Operating | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing | ||||||||
Storage | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing | ||||||||
Accessories |
|||||||||
Included | Operating Manual on CD; LMS XChange; Power Supply; Power Cord | Operating Manual on CD; AC or DC Power Input Connector Kit | Operating Manual on CD; AC or DC Power Input Connector Kit; Industrial RJ45 Kit; Fuse(s) |
||||||
Optional | Printed Operating Manual; REMOTE Display; REMOTE Display 10-foot Extension Cable; Network Adapter; Wireless Network Adapter; Isokinetic Sampling Probe; Sample Tubing; Data Cabling; Purge Filter; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; RS-232 Interface Cable and Adapter; High Pressure Diffuser (except 2014P); LMS Express, Express RT, RT+ and LMSNet software | Printed Operating Manual; Data Cable; Sample Tubing; 24VDC 100W Power Adapter (DC Model) |
OFFERTE AANVRAAG
The Remote 5102PN meets the stringent requirements of today's aseptic processing applications. Incorporating the latest in laser optical particle counting technology, Lighthouse designed the Remote 5102PN for continuous trouble-free operation.
The Remote 5102PN uses a built-in vacuum pump and flow control system to insure accurate and reliable sensor performance.
With a sensitivity of 0.5 micron at a controlled flow rate of 1.0CFM, the Remote 5102PN provides continuous real-time data at a cost-effective price per point.
The Remote 5102PN integrates seamlessly into large facility monitoring and management systems and transfers up to 2 channels of simultaneous particle count data using 4-20mA output.
Key Features:
- Channel sizes: 0.5 - 5.0 µm for GMP Compliance
- 1.0 CFM (2.83 LPM) Flow Rate
- Internal Vacuum Pump with Flow Control
- Up to 2 channels of simultaneous data
- Meets JIS Standards
- NEMA Style 316L Stainless Steel Enclosure or
- NEMA Style Powder-coated Steel Enclosure
- LED Status Indicators
- Interfaces to Facility Monitoring Systems
- 4-20mA Output
- 120 VAC or 24VDC, 100W Models Available
Type | Part # | Title | Paper Size | File Size | Download |
---|---|---|---|---|---|
Datasheet | - | Remote with Pump 5102PN | Letter | 509.7 KB | DOWNLOAD |
Operating Manual | 248083310-1 R3 | Remote 5012PN, 5102PN | Letter | 1,012.8 KB | DOWNLOAD |