NanoCount Liquid Particle Counters
TECHNICAL DETAILS
30 | 50 | 65C | |
Specifications |
|||
Size Range | 30 - 100nm | 50 - 200nm | 65 - 200nm |
Channel Sizes | Standard: 30nm, 50nm, 80nm, 100nm | Standard: 50nm, 100nm, 150nm, 200nm | Standard: 65nm, 100nm, 150nm, 200nm |
Flow Rate | 80 mL/min | 100 ml/min | |
Zero Count Level | < 20 counts/Liter | ||
Laser Source | Solid State Laser | Laser Diode | |
Data Storage | - | Rotating buffer, 2000 Records | |
NIST Traceable | |||
Communication Modes | Ethernet MODBUS TCP/IP, RS-232 via RJ45 to PC, RS-485 MODBUS | RS-485 and Ethernet use Industrial RJ45s and 4-20mA uses sealed DB9 | |
LED Indicators | - | Status | Power, Leak Alarm |
Display | TFT Color Touch Screen | ||
Concentration Limit | 1,000,000 counts/ml @ 5% Coincidence Error | 200,000 counts/ml @ 5% Coincidence Error | |
Enclosure | Stainless Steel | Polypropylene, White | |
Sample Inlet/Outlet Connection | 1/4" Flaretek™ | ||
Sample Temperature | 50 - 95°F (10 – 35°C) | 32 - 140°F (0 – 60°C) | 50 - 140°F (10 – 60°C) |
Sample Pressure | 75 PSI | ||
Wetted Surface Materials | Quartz, PFA, Kel-F, Kalrez, Ryton, Epoxy, Glass, Sapphire and SS316L | Quartz, PTFE, PFA, Viton, Kel-F, Kalrez, SS316 | Sapphire, PFA, Kel-F, Kalrez |
Communication Protocol | Ethernet, RS-485 MODBUS, four 4-20mA Analog Outputs | RS-485 MODBUS, 4-20mA, Ethernet TCP | |
Power | 100-240 VAC | 24VDC | |
Dimensions | 15.0" (w) x 12.0" (h) x 26.3" (d) [31.1 x 30.48 x 66.80 cm] | 9.00" (w) x 8.00" (h) x 11.00" (d) [22.86 x 20.32 x 27.94 cm] | |
Weight | 71 lbs (32.2 kg) | 11.8 lbs (32.2 kg) | 11.8 lbs (5.4 kg) |
Environmental Conditions |
|||
Operating | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing | ||
Storage | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing | ||
Accessories |
|||
Included | Operating Manual on CD; 10' Sample Tubing; Flow Cell Cleaning Solution and Brush; Cable; USB to RS485 Serial Adapter; LMS Express software | Operating Manual on CD; Connection Tubing; Flow Cell Cleaning Solution and Brush; Carrying Case; 2 each Industrial RJ45 kit; 24VDC Power Supply and Power Cord; Communication Cable | Operating Manual on CD; Connection Tubing; Flow Cell Cleaning Solution and Brush; 2 each Industiral RJ45 kit; 24VDC Power Supply and Power Cord |
Optional | Printed Operating Manual; Tubing | Printed Operating Manual; Remote Display; Remote Display 10-foot Extension cable; Additional Tubing; LMS Express, Express RT |
Chemicals that can be used with the NanoCount 65C
CHEMICAL | SYMBOL |
Water | H2O |
Deionized Water | H2O, Dl H2O |
Nitric Acid 70% | HNO3 70% |
Sulphuric Acid 96% | H2SO4 |
Hydrochloric Acid 37% | HCL |
Ammonium Hydroxide 29% | NH4OH |
Hydrogen Peroxide 30% | H2O2 |
Phosphoric Acid 86% | H3PO4 |
Hydrofluoric Acid 50% | HF 50% |
Hydrofluoric Acid 5% | HF 5% |
Hydrofluoric Acid 0.5% | HF 0.5% |
Tetramethylammonium Hydroxide | TMAH |
Potassium Hydroxide | KOH |
N-methyl Perryiodine | NMP |
Acetone | C3H6O |
Isopropyl Alcohol | IPA |
Chemicals that CAN NOT be used with the NanoCount 65C
CHEMICAL | SYMBOL |
Nitric Acid | HNO3 100% |
Ammonium Fluoride | NH4F |
Buffered Hydrofluoric Acid | BHF |
Buffered Oxide Etch | BOE |
GET A QUOTE
True Sensitivity
Second to None
Don't be fooled by sizing tricks, false count subtractions, or marketing gimmicks. When using in-situ particle counters for your UPW applications you want the most sensitive detection at the minimum particle size. The Nanocount line of UPW Liquid Particle Counters have the lowest false count rate, and are focused on true sensitivity.
Let us show you what you have been missing.