Home | Ultrapure Cleanroom Fogger
5012PN |
5102PN |
|||
Technical Specifications |
||||
Size Range | 0.5 - 25.0 μm | |||
Channel Sizes | Standard: 0.5, 5.0μm Optional: 0.5, 10.0μm |
|||
Counting Efficiency | 50% at the minimum detectable size | |||
Flow Rate | 0.1 CFM (2.83 LPM) | 1.0 CFM (28.3 LPM) | ||
Laser Source | Laser Diode | |||
Zero Count Level | <1 count / 5 minutes (per JIS) | |||
Concentration Limits | 2,000,000 particles / ft³ @ 5% coincidence loss | 500,000 particles / ft³ @ 5% coincidence loss | ||
Calibration | ISO 21051-4 compliant ISO 17025 Accredited | |||
Sample Output | Internally filtered to HEPA standards (>99.997% @ 0.3μm) | |||
Vacuum Requirements | Internal Vacuum Pump | |||
Air Flow | Internally Monitored and Controlled | |||
Communication Modes | 4-20mA Output | |||
LED Indicators | Power, Service | |||
Connector | 8-pin Phoenix | |||
External Pump Switch | 2-pin Phoenix dry-contact relay | |||
Enclosure | NEMA Style 316L Stainless Steel or Powder Coated Steel | |||
Power | 110 / 220 VAC, 50-60Hz or 24VDC 100W | |||
Dimensions | 10” (w) x 12” (h) x 6.9” (d) [25.4 x 30.4 x 14.9 cm] | |||
Weight | AC Model, 20 lbs (9 kg); DC Model, 19.5 lbs (8.85 kg) | |||
Environmental Conditions |
||||
Operating | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing | |||
Storage | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing | |||
Accessories |
||||
Included | Operating Manual on CD; AC or DC Power Input Connector Kit | |||
Optional | Printed Operating Manual; Data Cable; Sample Tubing; 24VDC 100W Power Adapter (DC Model) |
Portable Utrapure water Cleanroom fogger AFM24, for producing High-density Water fog for Air Flow visualization in Cleanrooms.
The new series foggers provides a high volume, density and purity of fog based on pure water. The water fog is generated by atomizing DI water into water droplets, which are nominally from 2 µm to 4 µm in size. The generated fog is ultrapure leaving no trace particles after its evaporation behind. The high density of the fog increases the duration and travel distance of the fog.
The Utrapure water Cleanroom fogger system is the only CE-certified system available on the market and IP 68 protected. The Utrapure Cleanroom fogger can be used in any class of cleanroom environment at rest and in operation.
Applications
Features
Standard system contains
Options
SERVICE & CALIBRATION