Remote 5104
Page 2

Remote 5104

Technical Details

  1104LD 2014 3014 5014 3104 5104 5104V

Specifications

Size Range 0.10 - 1.0 μm 0.2 - 2μm 0.3 - 25μm 0.5 - 25μm 0.3 - 25μm 0.5 - 25μm
Channel Sizes Standard: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0 μm Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Optional: 0.2, 0.3, 0.5, 1.0μm
Optional: 0.2, 0.3, 0.5, 2.0μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Optional: 0.3, 0.5, 1.0, 5.0μm
Standard: 0.5, 5.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm
Optional: 0.5, 5.0, 10.0, 25.0μm
Standard: 0.3, 0.5μm
Optional: 0.3, 0.5, 0.7, 1.0μm; 0.3, 0.5, 1.0, 5.0μm;
Optional: 0.3, 0.5, 5.0, 10.0μm
Standard: 0.5, 5.0μm; 0.5, 10.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm; 0.5, 5.0, 10.0, 25.0μm
Flow Rate 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM)
Laser Source Laser Diode
Zero Count Level < 1 Count / 5 minutes (meets specifications of JIS B 9921 and reporting requirements of ISO 21501-4 Annex C)
Concentration Limits 1,000,000 particles /ft3 (34,300,000/m3 ) @ 10% coincidence loss 4,000,000 Particles/ft3@10% Coincidence loss suitable for ISO classes 1-8 cleanrooms 1,000,000 Particles/ft3@10% Coincidence loss suitable for ISO classes 1-8 cleanrooms
Calibration NIST Traceable ISO 21501-4 Compliant and Accredited ISO 17025
Vacuum Requirements External Vacuum >18” (45.72 cm) of Hg
Air Flow Internally Monitored
Communication Modes Serial - 485 Modbus RJ45
LED Indicators Power, Flow, Service, Sampling
Data Storage Rotating Buffer, 2000 Records
Supporting Software LMS XChange, Lighthouse Monitoring System; LMS Express; Express RT; LMSNet
Environmental Sensors - (Optional) Temperature/Relative Humidity Probe 32 - 122F (0-50C)+/-1F (0.5C); 0 - 100% RH +/-2% -
Connector RJ45 Industrial RJ45
Power 24 VDC
Dimensions 10.5” (w) x 7.65” (h) x 3.35” (d) [26.67 x 19.43 x 8.51 cm] 1.7”(l) x 4.2”(w) x 3”(h) [4.3 x 10.7 x 7.6 cm] 1.7"(l) x 4.2"(w) x 3"(h) [4.3 x 10.7 x 7.6 cm] 1.7”(l)x4.2”(w)x3”(h) [4.3x10.7x7.6cm] 5.42” (l) x 2.14” (w) x 3.58” (h), [13.76 cm (l) x 5.43 cm (w) x 9.09 cm (h)]

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Get a Quote

Incorporating the latest laser optical particle counting technology, Lighthouse designed the Remote 5104 for continuous trouble-free operation. With a minimum sensitivity range of 0.5 µm and a flow rate of 1.0 CFM, the Remote 5104 provides continuous real-time data collection at a cost-effective price per point.

Using an external vacuum source, the Remote 5104 provides versatile mounting options and can be installed where space is at a premium. The Remote 5104 will integrate seamlessly into extensive facility monitoring and management systems.

Key Features:

  • 0.5 – 25 µm Particle Size Range
  • 1.0 CFM (28.3 LPM) Flow Rate
  • 4 Channels of Simultaneous Data
  • RS-485 Modbus Output
  • Meets 21501-4 Standards
  • Status Indicators
  • External Alarm Output
  • Stainless Steel Enclosure
  • Facility Monitoring Systems Interface
  • Versatile Mounting Options
  • External Vacuum Source
  • Designed for Reliability

Instrument Files

Title
Download
Operating Manual – Remote 2014, 3014, 5014, 5104 Gen B
Download
Datasheet – Remote 2014, 3014, 5014, 5104