Technical Details

  1104LD 2014 3014 5014 3104 5104 5104V

Specifications

Size Range 0.10 - 1.0 μm 0.2 - 2μm 0.3 - 25μm 0.5 - 25μm 0.3 - 25μm 0.5 - 25μm
Channel Sizes Standard: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0 μm Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Optional: 0.2, 0.3, 0.5, 1.0μm
Optional: 0.2, 0.3, 0.5, 2.0μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Optional: 0.3, 0.5, 1.0, 5.0μm
Standard: 0.5, 5.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm
Optional: 0.5, 5.0, 10.0, 25.0μm
Standard: 0.3, 0.5μm
Optional: 0.3, 0.5, 0.7, 1.0μm; 0.3, 0.5, 1.0, 5.0μm;
Optional: 0.3, 0.5, 5.0, 10.0μm
Standard: 0.5, 5.0μm; 0.5, 10.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm; 0.5, 5.0, 10.0, 25.0μm
Flow Rate 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM)
Laser Source Laser Diode
Zero Count Level < 1 Count / 5 minutes (meets specifications of JIS B 9921 and reporting requirements of ISO 21501-4 Annex C)
Concentration Limits 1,000,000 particles /ft3 (34,300,000/m3 ) @ 10% coincidence loss 4,000,000 Particles/ft3@10% Coincidence loss suitable for ISO classes 1-8 cleanrooms 1,000,000 Particles/ft3@10% Coincidence loss suitable for ISO classes 1-8 cleanrooms
Calibration NIST Traceable ISO 21501-4 Compliant and Accredited ISO 17025
Vacuum Requirements External Vacuum >18” (45.72 cm) of Hg
Air Flow Internally Monitored
Communication Modes Serial - 485 Modbus RJ45
LED Indicators Power, Flow, Service, Sampling
Data Storage Rotating Buffer, 2000 Records
Supporting Software LMS XChange, Lighthouse Monitoring System; LMS Express; Express RT; LMSNet
Environmental Sensors - (Optional) Temperature/Relative Humidity Probe 32 - 122F (0-50C)+/-1F (0.5C); 0 - 100% RH +/-2% -
Connector RJ45 Industrial RJ45
Power 24 VDC
Dimensions 10.5” (w) x 7.65” (h) x 3.35” (d) [26.67 x 19.43 x 8.51 cm] 1.7”(l) x 4.2”(w) x 3”(h) [4.3 x 10.7 x 7.6 cm] 1.7"(l) x 4.2"(w) x 3"(h) [4.3 x 10.7 x 7.6 cm] 1.7”(l)x4.2”(w)x3”(h) [4.3x10.7x7.6cm] 5.42” (l) x 2.14” (w) x 3.58” (h), [13.76 cm (l) x 5.43 cm (w) x 9.09 cm (h)]

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Get a Quote

Using state-of-the-art Long Life Laser Diode technology, the Lighthouse Remote 1104LD is an industry first- a laser diode-based remote particle counter with a sensitivity of 0.10 micron at a high flow rate of 1.0 CFM (28.3 LPM) that uses an optics and photodiode system designed to reduce or eliminate the temperature issues associated with high power laser diode sensors. Designed with reliability and quality in mind, the Remote 1104LD features our long-life laser diode technology that provides higher concentration limits, a better signal-to-noise ratio, and requires less routine maintenance than conventional HeNe laser-based particle counters. The Remote 1104LD supplies continuous real-time data at a cost-effective price per point. Very simple to integrate with large facility monitoring and management systems, the Remote 1104LD can transfer up to 8 channels of simultaneous particle count data via RS-485 MODBUS.

Key Features:

  • 0.10 – 5.0 µm Sensitivity
  • 1.0 CFM (28.3 LPM) Flow Rate
  • Up to 8 Channels of Simultaneous Data
  • World’s Smallest 1.0 CFM 0.10 µm counter
  • Meets ISO 21501-4 Standards
  • Long Life Laser Diode
  • Small Footprint
  • Status Indicators
  • RS-485 MODBUS (R1100 Pulse output)
  • Stainless Steel Enclosure
  • Interfaces to Facility Monitoring Systems
  • Uses External Vacuum Source
  • Designed for Reliability
  • Designed to Ensure 21 CFR Part 11 Compliance when used with Additional Software
  • No Routine Laser Tube Replacements

Instrument Files

Title
Download
Operating Manual – Remote 1100, 1104 Gen D
Download
Datasheet – Remote 1100,1104LD
Download