TECHNICAL DETAILS

  Remote 1100LD Remote 1104LD Remote 2010 Remote 3010 Remote 5010 Remote 2012 Remote 3012 Remote 5012 Remote 3102 Remote 5102 Remote 2014 Remote 3014 Remote 5014 Remote 3104 Remote 5104 Remote 5102V

Remote 5104V

Specifications

Size Range 0.10 - 1.0 μm 0.2 - 2.0 μm 0.3 - 25μm 0.5 - 25μm 0.2 - 2μm 0.3 - 25μm 0.5 - 25μm 0.3 - 25μm 0.5 - 25μm 0.2 - 2μm 0.3 - 25μm 0.5 - 25μm 0.3 - 25μm 0.5 - 25μm
Channel Sizes 2 Channels: 0.1, 0.3 μm
4 Channels: 0.1, 0.2, 0.3, 0.5 μm
Standard: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0 μm Standard: 0.2, 0.3 μm
Standard: 0.2, 0.5 μm
Standard: 0.3, 0.5 μm
Standard: 0.3, 1.0 μm
Standard: 0.3, 3.0 μm
Standard: 0.5, 5.0 μm
Optional: 0.5, 10.0 μm
Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Standard: 0.5, 5.0μm
Standard: 0.5, 10.0μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Standard: 0.5, 5.0μm
Standard: 0.5, 10.0μm
Standard: 0.2, 0.3μm
Standard: 0.2, 0.5μm
Optional: 0.2, 0.3, 0.5, 1.0μm
Optional: 0.2, 0.3, 0.5, 2.0μm
Standard: 0.3, 0.5μm
Standard: 0.3, 1.0μm
Standard: 0.3, 3.0μm
Optional: 0.3, 0.5, 1.0, 5.0μm
Standard: 0.5, 5.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm
Optional: 0.5, 5.0, 10.0, 25.0μm
Standard: 0.3, 0.5μm
Optional: 0.3, 0.5, 0.7, 1.0μm; 0.3, 0.5, 1.0, 5.0μm;
Optional: 0.3, 0.5, 5.0, 10.0μm
Standard: 0.5, 5.0μm; 0.5, 10.0μm
Optional: 0.5, 1.0, 5.0, 10.0μm; 0.5, 5.0, 10.0, 25.0μm
Standard: 0.5, 5.0μm;
Optional: 0.5, 10.0; 0.5, 25.0 μm
Standard: 0.5, 5.0 µm
Optional 4 Channels: 0.5, 1.0, 5.0, 25.0 µm
Counting Efficiency 50% @ 0.1 μm; 100% for particles > 0.15 μm (per JIS) 50% @ 0.2 μm; 100% for particles > 0.3 μm (per JIS) 50% @ 0.3 μm; 100% for particles > 0.45 μm (per JIS) 50% @ 0.5 μm; 100% for particles > 0.75 μm (per JIS) 50% @ 0.2μm; 100% for particles > 0.3μm (per JIS) 50% @ 0.3μm; 100% for particles > 0.45μm (per JIS) 50% @ 0.5μm; 100% for particles > 0.75μm (per JIS) 50% @ 0.3μm; 100% for particles > 0.45μm (per JIS) 50% @ 0.5μm; 100% for particles > 0.75μm (per JIS) 50% @ 0.2μm; 100% for particles > 0.3μm (per ISO 21501-4) 50% @ 0.3μm; 100% for particles > 0.45μm (per ISO 21501-4) 50% @ 0.5μm; 100% for particles > 0.75μm (per ISO 21501-4) 50% @ 0.3 μm; 100% for particles > 0.45 μm (per ISO 21501-4) 50% @ 0.5 μm; 100% for particles > 0.75 μm (per ISO 21501-4)
Flow Rate 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM)
Laser Source Long Life Laser Diode Laser Diode
Zero Count Level <1 count / 5 minutes (per JIS) < 1 count / 5 minutes (per ISO 21501-4)
Concentration Limits 500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss, per ISO 21501-4 2,000,000 Particles / ft3 @ 5% Coincidence Loss 500,000 Particles / ft3 @ 5% Coincidence Loss
Calibration NIST Traceable, per ISO 21501-4 NIST Traceable
Vacuum Requirements External Vacuum required, nominal 1.0 CFM >18" (45.72 cm) of Hg External Vacuum >18” (45.72 cm) of Hg
Air Flow - Internally Monitored Critical Orifice controlled Internally Monitored
Communication Modes Pulse Output via Single DB-15 Connector RS 485 MODBUS Pulse Output 4-20mA RS-485 MODBUS 4-20mA RS-485/MODBUS
LED Indicators Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling
Data Storage - Rotating Buffer, 2000 Records - Rotating Buffer, 2000 Records
Supporting Software Lighthouse Monitoring System, LMS Express, LMS Express RT, LMS Express RT Plus Lighthouse Monitoring System LMS XChange, Lighthouse Monitoring System; LMS Express; Express RT; LMSNet
Environmental Sensors - (Optional) Temperature/Relative Humidity Probe 32 - 122F (0-50C)+/-1F (0.5C); 0 - 100% RH +/-2% -
External Alarm Output - Normally Open (NO), Dry-contact, Rated 0-60 VDC/VAC, 1 Amp -
Connector Single DB-15 Connector   RJ45 (Data) Screw Terminal RJ45 Industrial RJ45
Enclosure Stainless steel 316L Stainless Steel, Wash Down Resistant 316L Stainless Steel, Wash Down Resistant/Splash-Tight
VHP Compatible - Yes
Power 24 VDC
Dimensions 10.5” (w) x 7.65” (h) x 3.35” (d) [26.67 x 19.43 x 8.51 cm] 1.7”(l) x 4.2”(w) x 2.5”(h) [4.3 x 10.7 x 6.3 cm] 3.5"(l) x 4.2"(w) x 3.3"(h) [8.89 x 10.66 x 8.38 cm] 1.7"(l) x 4.2"(w) x 3"(h) [4.3 x 10.7 x 7.6 cm] 5.4” (l) x 4.6” (h) x 1.65” (d) [13.72 x 11.68 x 4.19 cm] 4.2" (l) x 2.0" (w) x 3.7" (h) [10.7 x 5.08 x 9.398 cm]
Weight 7.0 lb (3.18 kg) 0.78 lb (0.35 kg) 0.97 lb (0.44 kg) 2 lb (0.90 kg) 0.97 lb (0.44 kg) 17.3 oz (0.49 kg) 15.6 oz (0.44 kg)

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Accessories

Included Operating Manual, Purge Filter, Power Supply, Power Cord Operating Manual on CD
Optional Isokinetic Sampling Probe, Sample Tubing, Cabling Printed Operating Manual; Isokinetic Sampling Probes; Vacuum Tubing; Sample Tubing; Cabling; Purge Filter; REMOTE Mounting Bracket. Printed Operating Manual; Isokinetic Sampling Probes; Vacuum Tubing; Sample Tubing; Data Cable; Purge Filter; Power Supply; High-Pressure Diffuser (not available for 2012); RS-232 Interface cable and adapter; USB to Serial Port Converter; REMOTE Mounting Bracket.  Printed Operating Manual; RS-485 Hub with Power Supply; RS-232 Communication Cable and Adapter; USB to Serial Port Converter; Network Adapter; Wireless Network Adapter; Isokinetic Sampling Probes; Vacuum Tubing; Sample Tubing; Data Cable; Purge Filter; Power Supply; Temperature/ Relative Humidity Probe; 8’ T/RH Probe Extension Cable; REMOTE Mounting Bracket; LMS XChange, LMS Express; Express RT, LMSNet software; RS-485 Communication Kit; High-Pressure Diffuser (not available for 2014) Isokinetic Sampling Probe, Sample Tubing, Cabling, Purge Filter, RS-485 Hub

GET A QUOTE

First Name (*)
Last Name (*)
Company Name (*)
Email (*)
Phone (*)
Address (*)
Message (*)
Security Code
Security Code ImageRefresh Button
(*)
(*) Required Fields
Remote 1104LD Image

Using state-of-the-art Long Life Laser Diode technology, the Lighthouse Remote 1104LD is an industry first- a laser diode based remote particle counter with a sensitivity of 0.10 micron at a high flow rate of 1.0 CFM (28.3 LPM) that uses an optics and photodiode system designed to reduce or eliminate the temperature issues associated with high power laser diode sensors. Designed with reliability and quality in mind, the Remote 1104LD features our long-life laser diode technology that provides higher concentration limits, a better signal to noise ratio and requires less routine maintenance than conventional HeNe laser-based particle counters. The Remote 1104LD supplies continuous real time data at a cost-effective price per point. Very simple to integrate with large facility monitoring and management systems, the Remote 1104LD can transfer up to 8 channels of simultaneous particle count data via RS-485 MODBUS.

Key Features:

  • 0.10 - 5.0 µm Sensitivity
  • 1.0 CFM (28.3 LPM) Flow Rate
  • Up to 8 Channels of Simultaneous Data
  • World’s Smallest 1.0 CFM 0.10 µm counter
  • Meets ISO 21501-4 Standards
  • Long Life Laser Diode
  • Small Footprint
  • Status Indicators
  • RS-485 MODBUS
  • Stainless Steel Enclosure
  • Interfaces to Facility Monitoring Systems
  • Uses External Vacuum Source
  • Designed for Reliability
  • Designed to Ensure 21 CFR Part 11 Compliance when used with Additional Software
  • No Routine Laser Tube Replacements
  • No Routine Laser Optics Cleaning
  • Cooler Running Dual-Avalanche Photodiode
  • Detector Array Design