Home | ApexR05
ApexR5 | ApexR02 | ApexR03 | ApexR05 | ||
Specifications |
|||||
Minimum Size Detection | 0.5 μm | 0.2 μm | 0.3 µm | 0.5 µm | |
Standard 2 channel sizes | 0.5, 5.0 μm | 0.2, 0.3 µm | 0.3, 0.5 µm | 0.5, 5.0 µm | |
Optional 4 channel sizes | 0.5, 1.0, 5.0, 10.0 μm | 0.2, 0.3, 0.5, 1.0 µm | 0.3, 0.5, 1.0, 5.0 µm | 0.5, 1.0, 5.0, 10.0 µm | |
Additional Available sizes | 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0 μm | 0.2, 0.3, 0.5, 0.7, 1.0, 2.0 µm | 0.3, 0.5, 0.7, 1.0, 3.0, 5.0 µm | 0.5, 0.7, 1.0, 3.0, 5.0, 7.0,10.0µm | |
Flow Rate / Nozzle Size | 1 CFM (28.3 LPM), 1/4” inlet and outlet fitting | 0.1 CFM (2.83 LPM), 1/8” inlet and outlet fitting | |||
Weight | 1.6 lbs (0.73 kg) | 1.4 lbs (0.63 kg) | |||
Communication | Ethernet or Serial | ||||
Modbus Protocol | TCP or RTU or ASCII | ||||
Data Storage Records | 3,000 | ||||
Self Diagnostics | Laser power supply, laser current, laser power, photo detector power supply, background voltage, photo detector health. | ||||
Web Server | Remote access via web browser | ||||
Validation Mode | Yes | ||||
Alarm Light | Yes | ||||
Enclosure | 304 Stainless Steel | ||||
Power | 24 VDC, PoE | ||||
External Vacuum Requirements | Nominal 1.0 CFM 18" (45.72 cm) of Hg | ||||
Light Source | Extreme Life Laser Diode | ||||
Calibration | ISO 21501-4 Compliant and Accredited ISO 17025 | ||||
Zero Count Level | < 1 Count / 5 minutes (meets specifications of JIS B 9921 and reporting requirements of ISO 21501-4 Annex C) | ||||
Concentration Limits 1.0 CFM | 1,000,000 particles/ft3@10% (34,300,000/m3) coincidence loss suitable for ISO classes 1-8 cleanrooms | ||||
Concentration Limits 0.1 CFM | 4,000,000 particles/ft3@10% (137,200,000/m3) coincidence loss suitable for ISO classes 1-8 cleanrooms | ||||
Dimensions | 6.38”x 2.93“x 4.63”(16.2x7.4x11.7cm) | ||||
Options |
|||||
Smart Bracket | Wall Mount Flush, Wall Mount Through Base, Wall Mount Through Wall |
Delivering the utmost data integrity, risk mitigation, and environmental compatibility, the ApexR05 Remote Particle Counter is unmatched in the industry. Designed with enhanced self-diagnostics for risk mitigation, the ApexR05 gives you the utmost confidence in your data. With its minimum detection of 0.5 μm and a long-life laser diode sensor, the ApexR05 provides unmatched performance, accuracy, and reliability.
The sloped front face and optional sealed connection provide you with compatibility for decontamination processes, including VHP. The ApexR05 stands alone in the industry and is truly the best in class.
Key Features:
SERVICE & CALIBRATION
The Apex smart bracket accessory contains the location instead of the particle counter, allowing the user to swap remotes quickly without making changes in the software, ensuring the location remains the same, reducing operator error, and improving location management
The Apex smart bracket accessory contains the location instead of the particle counter, allowing the user to swap remotes quickly without making changes in the software, ensuring the location remains the same, reducing operator error, and improving location management