NanoCount 65C
TECHNICAL DETAILS
30 | 50 | 65C | |
Specifications |
|||
Size Range | 30 - 100nm | 50 - 200nm | 65 - 200nm |
Channel Sizes | Standard: 30nm, 50nm, 80nm, 100nm | Standard: 50nm, 100nm, 150nm, 200nm | Standard: 65nm, 100nm, 150nm, 200nm |
Flow Rate | 80 mL/min | 100 ml/min | |
Zero Count Level | < 20 counts/Liter | ||
Laser Source | Solid State Laser | Laser Diode | |
Data Storage | - | Rotating buffer, 2000 Records | |
NIST Traceable | |||
Communication Modes | Ethernet MODBUS TCP/IP, RS-232 via RJ45 to PC, RS-485 MODBUS | RS-485 and Ethernet use Industrial RJ45s and 4-20mA uses sealed DB9 | |
LED Indicators | - | Status | Power, Leak Alarm |
Display | TFT Color Touch Screen | ||
Concentration Limit | 1,000,000 counts/ml @ 5% Coincidence Error | 200,000 counts/ml @ 5% Coincidence Error | |
Enclosure | Stainless Steel | Polypropylene, White | |
Sample Inlet/Outlet Connection | 1/4" Flaretek™ | ||
Sample Temperature | 50 - 95°F (10 – 35°C) | 32 - 140°F (0 – 60°C) | 50 - 140°F (10 – 60°C) |
Sample Pressure | 75 PSI | ||
Wetted Surface Materials | Quartz, PFA, Kel-F, Kalrez, Ryton, Epoxy, Glass, Sapphire and SS316L | Quartz, PTFE, PFA, Viton, Kel-F, Kalrez, SS316 | Sapphire, PFA, Kel-F, Kalrez |
Communication Protocol | Ethernet, RS-485 MODBUS, four 4-20mA Analog Outputs | RS-485 MODBUS, 4-20mA, Ethernet TCP | |
Power | 100-240 VAC | 24VDC | |
Dimensions | 15.0" (w) x 12.0" (h) x 26.3" (d) [31.1 x 30.48 x 66.80 cm] | 9.00" (w) x 8.00" (h) x 11.00" (d) [22.86 x 20.32 x 27.94 cm] | |
Weight | 71 lbs (32.2 kg) | 11.8 lbs (32.2 kg) | 11.8 lbs (5.4 kg) |
Environmental Conditions |
|||
Operating | 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing | ||
Storage | 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing | ||
Accessories |
|||
Included | Operating Manual on CD; 10' Sample Tubing; Flow Cell Cleaning Solution and Brush; Cable; USB to RS485 Serial Adapter; LMS Express software | Operating Manual on CD; Connection Tubing; Flow Cell Cleaning Solution and Brush; Carrying Case; 2 each Industrial RJ45 kit; 24VDC Power Supply and Power Cord; Communication Cable | Operating Manual on CD; Connection Tubing; Flow Cell Cleaning Solution and Brush; 2 each Industiral RJ45 kit; 24VDC Power Supply and Power Cord |
Optional | Printed Operating Manual; Tubing | Printed Operating Manual; Remote Display; Remote Display 10-foot Extension cable; Additional Tubing; LMS Express, Express RT |
Chemicals that can be used with the NanoCount 65C
CHEMICAL | SYMBOL |
Water | H2O |
Deionized Water | H2O, Dl H2O |
Nitric Acid 70% | HNO3 70% |
Sulphuric Acid 96% | H2SO4 |
Hydrochloric Acid 37% | HCL |
Ammonium Hydroxide 29% | NH4OH |
Hydrogen Peroxide 30% | H2O2 |
Phosphoric Acid 86% | H3PO4 |
Hydrofluoric Acid 50% | HF 50% |
Hydrofluoric Acid 5% | HF 5% |
Hydrofluoric Acid 0.5% | HF 0.5% |
Tetramethylammonium Hydroxide | TMAH |
Potassium Hydroxide | KOH |
N-methyl Perryiodine | NMP |
Acetone | C3H6O |
Isopropyl Alcohol | IPA |
Chemicals that CAN NOT be used with the NanoCount 65C
CHEMICAL | SYMBOL |
Nitric Acid | HNO3 100% |
Ammonium Fluoride | NH4F |
Buffered Hydrofluoric Acid | BHF |
Buffered Oxide Etch | BOE |
GET A QUOTE
65 nanometer liquid particle counter for chemicals.
NanoCount 50C and NanoCount 65C are two of the smallest 50nm and 65nm liquid particle counters on the market designed to monitor ultrapure process chemicals. Based on state-of-the-art technology and built by Lighthouse, they are backed by a two-year warranty and international support.
The NanoCount 50C and 65C have integrated manual flow control, built in leak detector and provides industry standard RS-485 Modbus and 4-20mA communication protocols so integration into facility monitoring systems or use as a stand-alone device is quick and easy.
Using its integrated Ethernet TCP/IP interface allows the NanoCount 50C and 65C to connect to standard corporate Ethernet LANs as easily as it connects to Facility Monitoring Systems.
The compact size of the NanoCount 50C and 65C allow them to be installed where other liquid particle counters won’t fit, including point-of-use process control applications.
Incorporating the latest in laser optical particle counting technology and designed for reliability, the NanoCount 50C and 65C are intended for continuous trouble-free operation.
Key Features:
- NanoCount 50C channel sizes: 50nm, 100nm, 150nm, 200nm.
- NanoCount 65C channel sizes: 65nm, 100nm, 150nm, 200nm.
- Four channels of simultaneous count data.
- 100 ml per minute flow rate.
- Ultra-pure process chemicals compatibility.
- Easily integrates with existing building automation and factory management systems.
- Industry standard RS-485/Modbus, 4-20mA and Ethernet interfaces (sealed connections).
- NEMA style enclosure.
- Built in leak detector
- Integrated manual flow control.
Type | Part # | Title | Paper Size | File Size | Download |
---|---|---|---|---|---|
Datasheet | 190906 | NanoCount65C | Letter | 938.6 KB | DOWNLOAD |