What are the differences between the different types of air particle counter calibrations?
Standard Size Calibration (includes as received data)
- The following three parameters will be tested and stated as per ISO 21501-4:2018
- Size Setting
- Sampling Flow Rate
- False count
- As received data will be stated for the following three parameters:
- Size Setting Error
- Sampling Flow Rate
- False count
ISO 21501-4 Calibration (includes as received data)
- The following five parameters will be tested and stated as per ISO 21501-4:2018
- Size Setting
- Counting Efficiency
- Sampling Flow Rate
- Size Resolution
- False count
- As received data will be stated for the following four parameters:
- Size Setting
- Sampling Flow Rate
- Size Resolution
- False count
(Reccomended) ISO 17025 Accredited Calibration (includes as received data) Click here to find out more What to expect from an ISO 17025 calibration
- The following five parameters will be tested and stated as per ISO/IEC 17025 and ISO 21501-4:2018 with as received data
- Size Setting
- Counting Efficiency
- Sampling Flow Rate
- Size Resolution
- False count
You can request a calibration on the RMA Website.