| Airborne Particle Counters |
| Remote Products > 5012 |
| REMOTE 5012 |
Incorporating the latest in laser optical particle counting
technology, Lighthouse designed the REMOTE 5012 for continuous trouble-free operation.
With a sensitivity of 0.5, at a
flow rate of 0.1 CFM (2.83 LPM), the REMOTE 5012 provide real-time continuous data collection at a
cost-effective price per point.
Using an external vacuum source, this compact particle
counters provide versatile mounting options and can be
installed where space is at a premium.
The REMOTE5012 integrate seamlessly
into large facility monitoring and management systems and
transfer up to 2 channels of simultaneous particle count data
using 4-20 mA output.
Designed and built by Lighthouse, a name you can trust. |

[ Virtual Demo ] |
|
| |
 |
 |
 |
 |
 |
 |
| Online Monitoring |
Isokinetic
Probe |
Screw Terminal
Connector |
Stainless Steel
Enclosure |
LED Status |
Barbed Vacuum Connector |
|
|
| Features |
Benefits |
- 0.5 - 25μm Sensitivity
- 0.1 CFM (2.83 LPM) Flow Rate
- Up to 2 Channels of Simultaneous Data
- All Models Meet JIS Standards
- Compact Size
- Stainless Steel Enclosure
- Status Indicators
- Alarm Output
- 4-20 mA Output
- Facility Monitoring Systems Interface
- Versatile Mounting Options
- External Vacuum Source
- Designed for Reliability
|
- 2 Year Warranty
- Install in Limited Space
- Cost Effective Price per Point
- Proven Data Integrity
- Proven Reliability
- Continuous Real-Time Data
- Easy System Integration
- International Support
- Low Cost of Ownership
|
|
|
| |
| Specifications |
Size Range:
|
0.5 - 25μm |
Channel Sizes:
|
Standard: 0.5, 5.0μm
Standard: 0.5, 10.0μm |
Counting Efficiency:
|
50% @ 0.5μm; 100% for particles
> 0.75μm (per JIS) |
Flow Rate:
|
0.1 CFM (2.83 LPM) |
Laser Source:
|
Laser Diode |
Zero Count Level:
|
<1 count / 5 minutes (per JIS) |
Concentration Limits:
|
2,000,000 Particles / ft3 @ 5% Coincidence Loss |
Calibration:
|
NIST Traceable |
Vacuum Requirements:
|
External Vacuum >18” (45.72 cm) of Hg |
Communication Modes:
|
4-20mA |
LED Indicators:
|
Power, Service |
Connector:
|
Screw Terminal |
Supporting Software:
|
Lighthouse Monitoring System |
External Alarm Output::
|
Normally Open (NO), Dry-contact, Rated 0-60 VDC/VAC, 1 Amp |
Enclosure:
|
Stainless steel |
| Power: |
24 VDC |
Dimensions:
|
1.7”(l) x 4.2”(w) x 2.5”(h) [4.3 x 10.7 x 6.3 cm] |
| Weight: |
0.78 lb (0.35 kg) |
| Environmental Conditions |
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included: |
Operating Manual on CD |
| Optional: |
Printed Operating Manual; Isokinetic Sampling Probes; Vacuum Tubing; Sample Tubing; Data Cable;
Purge Filter; Power Supply; High-Pressure Diffuser (not available for 2012); RS-232 Interface cable
and adapter; USB to Serial Port Converter; REMOTE Mounting Bracket. |
| Sample System Integration |
 |
|
|
| |
| [ Back to top ] |
|
|
|
|
| |
| |
|