| Specifications |
Size Range:
|
0.2 - 2.0 μm |
Channel Sizes:
|
Standard: 0.2, 0.3, 0.5, 1.0 μm |
Counting Efficiency:
|
50% @ 0.2 μm; 100% for particles
> 0.3 μm (per JIS) |
Flow Rate:
|
0.1 CFM (2.83 LPM) |
Laser Source:
|
Laser Diode |
Zero Count Level:
|
<1 count / 5 minutes (per JIS) |
Concentration Limits:
|
2,000,000 Particles / ft3 @ 5% Coincidence Loss |
Calibration:
|
NIST Traceable |
Vacuum Requirements:
|
Internal Vacuum Pump |
Air Flow:
|
Internally Monitored and Controlled |
Communication Modes:
|
RS-485 MODBUS |
LED Indicators:
|
Power, Flow, Service, Sampling |
Data Storage:
|
3,000 Samples |
Connectors:
|
RJ45 |
Supporting Software:
|
LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and RT+; LMSNet |
| Up to Four 4-20 mA Inputs: |
Can be used with the Lighthouse standard 4-20 mA Temperature/Relative Humidity, Air Velocity or Differential
Pressure Probes or any other 4-20 mA device |
External Alarm Output:
|
Normally Open Dry Contact Rated 0-60 V AC/DC 1 Amp |
Enclosure:
|
Stainless Steel |
| Power: |
24 VDC |
Dimensions:
|
9”(l) x 3.5”(w) x 6”(h) [22.86 x 8.89 x 15.24 cm] |
| Weight: |
3.19 lbs (1.45 kg) |
| Environmental Conditions |
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included: |
Operating Manual on CD; LMS XChange; Power Supply; Power Cord |
| Optional: |
Printed Operating Manual; REMOTE Display; REMOTE Display 10-foot Extension Cable; Network Adapter;
Wireless Network Adapter; Isokinetic Sampling Probe; Sample Tubing; Data Cabling; Purge Filter; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; RS-232 Interface Cable and Adapter; LMS Express, Express RT, RT+ and LMSNet software |
| Sample System Integration |
 |