| Airborne Particle Counters |
| Portable Products > SOLAIR 5010+ |
| SOLAIR 5010+ |
| |
Solair 5010+ Portable Particle Counter
The SOLAIR 5010+ line of portable counters offers a sensitivity of 0.3 micron and a flow rate of 0.1 CFM (2.83 LPM) with a wide dynamic range up to 25.0 microns.
Utilizing a large, color 5.7” (14.47 cm) touch screen
interface, the SOLAIR 5010+ is easy to
configure and operate. The SOLAIR 5010+ can store a large amount of particle count data from 6
channels and data from up to 4 environmental sensors (4-20
mA). All data can be quickly and reliably downloaded to
a computer or printed to its built-in thermal printer.
Demonstrating its versatility, the SOLAIR 5010+ can be used as a portable instrument or integrated with a larger facility monitoring system.
Designed and built by Lighthouse Worldwide Solutions, a name you can trust.
|

[ Virtual Demo ] |
|
| |
 |
 |
 |
 |
 |
 |
| New Easy Loading Thermal Printer |
User Removable Rechargeable Li-Ion Battery |
Zoom Screen
Functionality |
Alpha Numeric
Location Labels |
User Friendly
Configuration |
Multiple
Cleanroom
Standard Reports
via Printer |
|
|
| Features |
Benefits |
- 0.5 µm Sensitivity
- 0.1 CFM (2.83 LPM) Flow Rate
- View 6 Particle Sizes Simultaneously
- Wide Dynamic Range, 0.3 to 25.0 µm
- Meets JIS Standards
- 5.7” (14.47 cm) Color Touch Screen Interface
- Removable/Rechargeable Li-Ion Battery
- Large Memory for Storing Data (3,000 samples)
- Up to 4 Environmental Sensors (4-20 mA)
- Alphanumeric Location Labels
- Easy to Load Thermal Printer
- User Definable Printouts
- Data Reported in Normalized Counts, FS-209E, ISO-14644-1,
BS-5295 or EC GMP
- Easily Configurable Interface with Zoom Capability
- Internal Audible Alarm
- External Alarm Trigger
- Data Easily Downloaded to Computer
|
- 2 Year Warranty
- User Friendly Interface
- Easy-to-Read Display
- International Sales and Service Support
- Low Cost of Ownership
|
|
|
| |
| Specifications |
Size Range:
|
0.5 - 25.0 µm |
Channel Sizes:
|
Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 µm
Standard: 0.5, 1.0, 3.0, 5.0,10.0, 25.0 µm |
Counting Efficiency:
|
50% @ 0.5µm; 100% for particles > 0.75µm
(per JIS) |
Flow Rate:
|
0.1 CFM (2.83 LPM) |
Laser Source:
|
Laser Diode |
Zero Count Level:
|
<1 count / 5 minutes (per JIS) |
Calibration:
|
NIST Traceable |
Count Modes:
|
Automatic, Manual, Beep, Concentration, Cumulative/Differential |
Data Storage:
|
3,000 sample records, rotating buffer (includes particle
and environmental
data, plus location and time) |
Communication Modes:
|
RS-232 via RJ-45 to PC
RS-485 MODBUS |
Supporting Software:
|
LMS XChange, LMS Express, LMS Express RT and LMS Express
RT+, LMS Net |
| Environmental Sensors: |
Optional (up to 4): Temperature, Relative Humidity,
Air Velocity, Differential Pressure or others |
| Concentration Limits: |
4,000,000 particles / ft3 @ 5% Coincidence
Loss |
Touch Screen Display:
|
5.7” (14.47 cm) Color, 320x240 |
| Printer: |
Built-in Thermal Printer |
| Reports: |
FS-209E, ISO-14644-1, BS-5295 & EC GMP |
| Alarm: |
Internal, Adjustable Alarm Buzzer, External Alarm Connector.
Alarms on Counts, Low Battery, Sensor Failure |
| Enclosure: |
Stainless steel |
| Sample Output: |
Internally filtered to HEPA standards (>99.997% @
0.3 µm) |
Vacuum Source:
|
Internal Pump, Automatic Flow Control |
| Power: |
85-264 VAC, 47-63 Hz |
Battery:
|
Li-Ion, Removable & Rechargeable |
Dimensions:
|
8” (l) x 8.5” (w) x 8.9” (h) [20.32
x 21.59 x 22.60 cm] |
| Weight: |
With Battery: 14.75 lbs (6.7 kg) Without
Battery: 12.75 lbs (5.8 kg) |
| Environmental Conditions |
| Operating: |
50°F to 104°F (10°C to 40°C)
/ 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C)
/ Up to 98% non-condensing |
| Accessories |
| Included: |
Operating Manual on CD, Isokinetic Sample
Probe with Tripod and Tubing, Purge Filter, Removable/Rechargeable
Li-Ion Battery, Power Cable and Spare Fuse, LMS XChange
Software, Printer, Printer Paper |
| Optional: |
Shipping Case with Wheels, Cleanroom Cart,
Temperature/Relative Humidity Probe, Air Velocity Probe,
Differential Pressure Probe, Filter Scanning Probe, High
Pressure Diffuser, Spare Removable/Rechargeable Li-Ion
Battery, External Battery Charger, LMS Express Software,
LMS Express RT Software, LMS Express Software RT+, Validation
Documentation |
|
|
| |
| [ Back to top ] |
|
|
|
|
| |
| |
|