| Specifications |
Size Range:
|
0.3 - 25.0μm |
Channel Sizes:
|
0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm
0.3, 0.5, 1.0, 5.0, 10.0, 25.0μm |
Flow Rate:
|
100 LPM or 3.53 CFM |
Counting Efficiency:
|
3350 = 50% @ 0.3μm; 100% for particles > 0.45μm, per ISO 21501-4 |
| Light Source: |
Extreme Life Laser Diode Technology with >20 Year Mean Time To Failure (MTTF) |
Zero Count Level:
|
<1 count / 5 minutes (per JIS) |
| Concentration Limits: |
200,000 particles / ft3 (7,060,00 particles / m3) @ 5% coincidence loss. Suitable for all Cleanrooms ISO Classes 1 thru 9 |
Calibration:
|
Factory Calibrated to ISO 21501-4 including Resolution and Counting Efficiency tests using NIST-traceable particles |
Count Modes:
|
Manual, Automatic, Concentration and Real Time (with LMS Pharma Edition or LMS Express 7.0 RT) |
Data Storage:
|
3,000 Sample Records (Rotating Buffer) which includes particle data, environmental data, locations and times |
Communication Modes:
|
RS-485/MODBUS; RS-232; Ethernet TCP/IP; Universal Serial Bus; USB Flash Drive |
Supporting Software:
|
LMS XChange, LMS Express 7.0, LMS Express RT, LMS Pharma Edition, ANY OPC Client (with LMS Express 7.0 OPC Server) |
Touch Screen Display:
|
5.7-inch (14.47 cm) color, 320x240 |
| Printer: |
Built-in Thermal Printer |
| Reports: |
ISO-14644-1, FS-209E & EU GMP Annex 1 |
| Alarm: |
Internal Adjustable Alarm, Alarms on Counts and Low Battery |
Enclosure:
|
304 Stainless Steel |
| Sample Output: |
Internally filtered to HEPA standards (>99.997% @ 0.3μm) |
Vacuum Source:
|
Internal closed-loop flow-controlled pump |
| Power: |
100-240 VAC, 50-60 Hz |
| Battery: |
Li-Ion, Removable & Rechargeable: Typical Battery Operation - *4 hours; Typical Re-charge Time - 4 hours |
Dimensions:
|
14.77” (d) x 8.0” (w) x 10.41” (h) [37.52 x 20.32 x 26.44 cm] |
Weight:
|
With Battery: 25.0 lbs (11.34 kg) Without Battery: 22.0 lbs (9.98 kg) |
| |
* typical battery operation of 10-minute sample, 10-minute hold time, repeat |
| Environmental Conditions |
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included: |
Operating Manual on CD; Data Cable; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Removable / Rechargeable Li-Ion Battery; Power Cable and Spare Fuse; LMS XChange Software; Printer Paper |
| Optional: |
Printed Operating Manual; Ethernet Network Adapter; Wireless Network Adapter; 4-20mA Temperature/Relative Humidity Probe or Wand; Cleanroom Cart; Cleanroom Thermal Printer Paper; Spare Removable/Rechargeable Li-Ion Battery; External Battery Charger; USB Flash Drive; LMS Express RT Software; LMS Pharma Edition Software; Validation Manual |