| Airborne Particle Counters |
| Portable Products > SOLAIR 3200+ |
| SOLAIR 3200+ |
| |
SOLAIR 3200+ Airborne Particle Counter
The SOLAIR 3200+ portable counter
offer a sensitivity of 0.3 micron at a flow rate of 2.0 CFM (56.6 LPM) or 50 LPM and a wide
dynamic range of up to 25.0 microns.
Utilizing a 5.7-inch color (14.5 cm) touch screen interface, the SOLAIR 3200+ is easy to configure
and operate. The SOLAIR 3200+ can store 3,000 records of particle count
data from up to 8 channels and environmental data from up
to four analog sensors (4-20 mA). All data can be quickly
and reliably transferred to a computer or printed on its built-in
thermal printer.
Demonstrating their versatility, the SOLAIR 3200+ can be used as portable instruments or
integrated into a facility monitoring system.
Designed and built by Lighthouse Worldwide Solutions, a name you can trust. |

[ Virtual Demo ] |
|
| |
 |
 |
 |
 |
 |
 |
| Easy Loading Thermal Printer |
User Removable
Rechargeable
Li-Ion Battery |
Zoom Screen
Functionality |
User Friendly
Configuration |
Alpha Numeric
Location Labels |
Multiple
Cleanroom
Standard Reports
via Printer |
|
|
| Features |
Benefits |
- 0.3μm to 25.0μm Range
- 2.0 CFM or 50 LPM Flow Rate
- Simultaneous Display of up to 8 Particle Sizes
- Wide Dynamic Range
- Both Models Meet JIS Standards
- 5.7-inch (14.5 cm) Color Touch Screen
- Removable, Rechargeable Li-Ion Battery
- 3,000-sample Data Storage Memory
- Up to 4 Environmental Sensors (4-20 mA)
- 200 Alphanumeric Location Labels
- Easy-to-Load Thermal Printer
- User-Definable Printouts
- Data Reported in Normalized Counts:
- FS-209E, ISO-14644-1, BS-5295 or EU GMP
- Easily Configured Interface with Zoom Capability
- Integrated Alarm and External Alarm Connection
- Data Easily Transferred to a Computer
|
- 2 Year Warranty
- User Friendly Interface
- Easy-to-Read Display
- International Support
- Low Cost of Ownership
|
|
|
| |
| Specifications |
Size Range:
|
0.3 - 25.0μm |
Channel Sizes:
|
0.3, 0.5,1.0, 3.0, 5.0, 10.0μm
0.3, 0.5,1.0, 5.0, 10.0, 25.0μm
0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0μm
0.3, 0.5, 0.7, 1.0, 5.0, 7.0, 10.0, 25.0μm |
Counting Efficiency:
|
50% @ 0.3μm; 100% for particles > 0.45μm (per JIS) |
Flow Rate:
|
2.0 CFM or 50 LPM |
Laser Source:
|
Laser Diode |
Zero Count Level:
|
<1 count / 5 minutes (per JIS) |
| Concentration Limits: |
250,000 Particles/ft3 @5% Coincidence Loss |
Calibration:
|
NIST Traceable |
Count Modes:
|
Automatic, Manual, Beep, Concentration, Cumulative/Differential and Real Time (with LMSNet or LMS Express RT/RT+) |
Data Storage:
|
3,000 Sample Records (Rotating Buffer) which includes particle and environmental data with locations and times |
Communication Modes:
|
RS-232 via RJ-45 to PC, RS-485/MODBUS |
Supporting Software:
|
LMS XChange, LMS Express, LMS Express RT and LMS Express RT+, LMSNet |
| Environmental Sensors: |
Optional (up to 4): Temperature/Relative Humidity, Air Velocity, Differential Pressure, or others |
Touch Screen Display:
|
5.7-inch (14.5 cm) Color, 320x240 |
| Printer: |
Built-in Thermal Printer |
| Reports: |
FS-209E, ISO-14644-1, BS-5295 & EU GMP |
| Alarm: |
Integrated, Adjustable Alarm Buzzer, External Alarm Connector. Alarms on Counts, Low Battery, Sensor Failure |
Enclosure:
|
Stainless Steel |
| Sample Output: |
Internally Filtered to HEPA Standards (>99.997% @ 0.3μm) |
Vacuum Source:
|
Internal Pump, Automatic Flow Control |
| Power: |
85-264 VAC, 47-63 Hz |
Battery:
|
Li-Ion, Removable & Rechargeable |
Dimensions:
|
14.65” (l) x 8” (w) x 10.4” (h) [37 x 20 x 26 cm] |
| Weight: |
With Battery: 26 lbs (11.8 kg) Without Battery: 23 lbs (10.43 kg) |
| Environmental Conditions |
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included: |
Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Removable
Rechargeable Li-Ion Battery; Power Cable and Spare Fuse; LMS XChange Software; Printer; Printer Paper |
| Optional: |
Printed Operating Manual; Network Adapter; Wireless Network Adapter; Carrying Case with Wheels;
Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter
Scanning Probe; Spare Removable Rechargeable Li-Ion Battery; External Battery Charger; LMS Express, RT,
RT+ Software; LMSNet Software; Validation Documentation |
|
|
| |