| Airborne Particle Counters |
| Portable Products > SOLAIR 3100+ |
| SOLAIR 3100+ |
| |
SOLAIR 3100+ Airborne Particle Counter
Dependable and reliable, the Lighthouse SOLAIR 3100+ laser diode particle counter use the latest in
optical particle counting technology.
The SOLAIR 3100+ provides
excellent sensitivity of 0.3, at a flow
rate of 1.0 CFM (28.3 LPM) and a dynamic range of up to
25.0μm.
The integrated 5.7-inch (14.47 cm) color touch screen interface
makes these instruments easy to configure and operate. Both
counters can store 3,000 records of particle count data from
up to 8 channels and environmental data from up to four
analog sensors (4-20mA). All data can be quickly and reliably
transferred to a computer or printed on its built-in thermal
printer.
Demonstrating their versatility, the SOLAIR 3100+ can be used as a portable instrument or integrated into
a facility monitoring system.
Designed and built by Lighthouse Worldwide Solutions, a name you can trust. |

[ Virtual Demo ]
|
|
| |
 |
 |
 |
 |
 |
 |
Easy Loading
Thermal Printer |
User Removable
Rechargeable Li-Ion Battery |
Zoom Screen
Functionality |
User Friendly
Configuration |
Alpha Numeric
Location Labels |
Multiple
Cleanroom
Standard Reports
via Printer |
|
|
| Features |
Benefits |
Applicaitons |
- 0.3μm - 25.0μm Range
- 1.0 CFM (28.3 LPM) Flow Rate
- User Removable, Rechargeable Li-Ion Battery
- Simultaneous Display of up to 8 Particle Sizes
- Both Models Meet JIS Standards
- Stainless Steel Enclosure
- Up to 4 Environmental Sensors (4-20mA)
- 5.7-inch (14.47 cm) Color Touch Screen
- Easy-to-Load Thermal Printer
- User-Defi nable Printouts
- 200 Alphanumeric Location Labels
- 3000-record Data Storage Memory
- Data Reported in Normalized Counts: FS-209E, ISO-14644-1, BS-5295 or EU GMP
- Integrated Alarm and External Alarm Connection
- Data Easily Transferred to a Computer
|
- 2 Year Warranty
- Small Footprint
- International Support
- Low Cost of Ownership
|
- Cleanroom Monitoring
- Facility Certification
- Facility Monitoring
- Disk Drive Industry
- Spray Paint Booths
- Cleanroom Packaging
- Aerospace
- Hospital Surgical Rooms
- Pharmaceutical
- Medical Device Assembly
|
|
|
| |
| Specifications |
Size Range:
|
0.3 - 25.0μm |
Channel Sizes:
|
0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm
0.3, 0.5, 1.0, 5.0, 10.0, 25.0μm |
| 8-Channel: |
0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0μm |
Flow Rate:
|
1.0 CFM (28.3 LPM) |
Counting Efficiency:
|
50% @ 0.3μm; 100% for particles > 0.45μm (per JIS) |
Laser Source:
|
Laser Diode |
Zero Count Level:
|
<1 count / 5 minutes (per JIS) |
| Concentration Limits: |
500,000 particles / ft3 @ 5% coincidence loss |
Calibration:
|
NIST Traceable |
Count Modes:
|
Manual, Automatic, Beep, Concentration and Real Time (with LMSNet or LMS Express RT/RT+) |
Data Storage:
|
3,000 Sample Records (Rotating Buffer) which includes particle and environmental data, locations and times |
Communication Modes:
|
RS-232 via RJ-45 to PC, RS-485/MODBUS |
Supporting Software:
|
LMS XChange, LMS Express, LMS Express RT/RT+, LMSNet |
| Environmental Sensors: |
Optional (up to 4): Temperature/Relative Humidity, Air Velocity, Differential Pressure |
Touch Screen Display:
|
5.7-inch (14.47 cm) color, 320x240 |
| Printer: |
Built-in Thermal Printer |
| Reports: |
ISO-14644-1, FS-209E, BS-5295 & EU GMP |
| Alarm: |
Integrated, Adjustable Alarm and External Alarm Connection: Alarms on Counts and Low Battery |
Enclosure:
|
Stainless Steel |
| Sample Output: |
Internally filtered to HEPA standards (>99.997% @ 0.3μm) |
Vacuum Source:
|
Internal flow-controlled pump |
| Power: |
100-240 VAC, 50-60 Hz |
| Battery: |
Li-Ion, Removable & Rechargeable |
Dimensions:
|
9.96” (l) x 8.0” (w) x 10.16” (h) [25.3 x 20.32 x 25.81 cm] |
| Weight: |
With Battery: 14.9 lbs (6.8 kg); Without Battery: 12.95 lbs (5.9 kg)
|
| Environmental Conditions |
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
| Accessories |
| Included: |
Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Removable/Rechargeable
Li-Ion Battery; Power Cable and Spare Fuse; LMS XChange Software; Printer; Printer Paper |
| Optional: |
Printed Operating Manual; Network Adapter; Wireless Network Adapter; Shipping Case with Wheels; Cleanroom
Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe;
High Pressure Diffuser; Spare Removable/Rechargeable Li-Ion Battery; External Battery Charger; 32-port Universal
Manifold System; LMS Express/RT/RT+ Software; LMSNet Software; Validation Documentation |
|
|
| |
| [ Back to top ] |
|
|
|
|
| |
| |
|