TECHNICAL DETAILS

  Remote 2014P Remote 3014P Remote 5014P Remote 3104P Remote 5104P Remote 5012PN Remote 5014PN Remote 5102PN Remoite 5104PN

Specifications

Size Range 0.2 - 2.0 μm 0.3 - 25.0 μm  0.5 - 25.0 μm 0.3 - 25.0 μm 0.5 - 25.0 μm
Channel Sizes Standard: 0.2, 0.3, 0.5, 1.0 μm Standard: 0.3, 0.5, 1.0, 5.0 μm Standard: 0.5, 1.0, 5.0, 10.0 μm Standard: 0.3, 0.5, 1.0, 5.0 μm Standard: 0.5, 1.0, 5.0, 10.0 μm Standard: 0.5, 5.0μm
Optional: 0.5, 10.0μm
Counting Efficiency 50% @ 0.2 μm; 100% for particles
> 0.3 μm (per ISO 21501-4)
50% @ 0.3 μm; 100% for particles
> 0.45 μm (per ISO 21501-4)
50% @ 0.5μm; 100% for particles > 0.75μm (per ISO 21501-4) 50% @ 0.3 μm; 100% for particles > 0.45 μm (per
JIS)
50% @ 0.5 μm; 100% for particles > 0.75 μm (per JIS) 50% @ 0.5μm, 100% for particles >0.75μm (per JIS) 50% @ 0.5μm, 100% for particles >0.75μm (per ISO 21501-4) 50% @ 0.5μm, 100% for particles >0.75μm (per JIS) 50% @ 0.5μm, 100% for particles >0.75μm (per ISO 21501-4)
Flow Rate 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM)
Laser Source Laser Diode
Zero Count Level <1 count / 5 minutes (per ISO 21501-4) <1 count / 5 minutes (per JIS); Zero Count with Vibration: 0.02g2/Hz from 8-66 Hz <1 count / 5 minutes (per JIS)
Concentration Limits 2,000,000 Particles / ft3 @ 5% Coincidence Loss 500,000 Particles / ft3 @ 5% Coincidence Loss 2,000,000 particles / ft³ @ 5% coincidence loss 500,000 particles / ft³ @ 5% coincidence loss
Calibration Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter NIST Traceable
Sample Output - Internally filtered to HEPA standards (>99.997% @ 0.3μm)
Vacuum Requirements Internal Vacuum Pump
Air Flow Internally Monitored and Controlled
Communication Modes RS-485 MODBUS, Ethernet, USB Ethernet, RS-485 MODBUS, Serial USB 4-20mA Output MODBUS: RS-485; Optional Ethernet MODBUS TCP 4-20mA Output MODBUS: RS-485; Optional Ethernet MODBUS TCP
LED Indicators Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling
Data Storage 3,000 Samples - Rotating Buffer, 2.000 Records - Rotating Buffer, 2.000 Records
Connector RJ45, USB (Type B) 2 RJ45, 2 RJ12 and one USB ‘B’ 8-pin Phoenix RJ45
External Pump Switch - 2-pin Phoenix dry-contact relay - -
Supporting Software LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and LMSNet  LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and RT+; LMSNet  Lighthouse Monitoring System; LMS Xchange Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT Lighthouse Monitoring System; LMS Xchange Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT
Up to Four 4-20 mA Inputs Can be used with the Lighthouse standard 4-20 mA Temperature/Relative Humidity, Air Velocity or Differential Pressure Probes or any other 4-20 mA device -
External Alarm Output - Normally Open Dry Contact Rated 0-60 V AC/DC 1 Amp -
Enclosure Stainless Steel NEMA Style 316L Stainless Steel or Powder Coated Steel
Power 24 VDC 110 / 220 VAC, 50-60Hz or 24VDC 100W
Dimensions 7.25”(l) x 3.75”(w) x 6”(h) [18.41 x 9.52 x 15.24 cm]  9.5”(l) x 6.5”(w) x 8.51”(h) [24.13 x 16.51 x 21.62 cm]  10” (w) x 12” (h) x 6.9” (d) [25.4 x 30.4 x 14.9 cm]
Weight 5.0 lbs (2.26 kg) 12.33 lbs (5.59 kg) AC Model, 20 lbs (9 kg); DC Model, 19.5 lbs (8.85 kg)

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Accessories

Included Operating Manual on CD; LMS XChange; Power Supply; Power Cord     Operating Manual on CD; AC or DC Power Input Connector Kit Operating Manual on CD; AC or DC Power Input Connector Kit;
Industrial RJ45 Kit; Fuse(s) 
Optional  Printed Operating Manual; REMOTE Display; REMOTE Display 10-foot Extension Cable; Network Adapter; Wireless Network Adapter; Isokinetic Sampling Probe; Sample Tubing; Data Cabling; Purge Filter; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; RS-232 Interface Cable and Adapter; High Pressure Diffuser (except 2014P); LMS Express, Express RT, RT+ and LMSNet software Printed Operating Manual; Data Cable; Sample Tubing; 24VDC 100W Power Adapter (DC Model) 

GET A QUOTE

First Name (*)
Last Name (*)
Company Name (*)
Email (*)
Phone (*)
Address (*)
(*) Required Fields
Message (*)
Security Code
Security Code ImageRefresh Button
(*)
Remote 5104PN Image

The Remote 5104PN meets the stringent requirements of today’s aseptic processing applications. Incorporating the latest laser optical particle counting technology, Lighthouse designed the Remote 5104PN for continuous trouble-free operation.

The  Remote 5104PN uses a built-in vacuum pump and flow control system to insure accurate and reliable sensor performance.

With a sensitivity of 0.5 micron at a controlled flow rate of 1.0 CFM, the  Remote 5104PN provides continuous real-time data at a cost-effective price per point.

The  Remote 5104PN integrates seamlessly into large facility monitoring and management systems and transfers up to 2 channels of simultaneous particle count data using RS-485 MODBUS or Ethernet MODBUS TCP.

 

Key Features:

·       Channel sizes: 0.5 - 5.0μm for GMP Compliance

·       1.0 CFM (28.3 LPM) Flow Rate

·       Internal Vacuum Pump with Flow Control

·       Up to 2 Channels of Simultaneous Data

·       Meets ISO 21501-4 Standards

·       NEMA Style 316L Stainless Steel Enclosure or

·       NEMA Style Powder Coated Steel Enclosure

·       LED Status Indicators

·       Interfaces to Facility Monitoring Systems

·       RS-485 MODBUS ASCII or Ethernet MODBUS TCP Output

·       120VAC or 24VDC, 100W Models Available