TECHNICAL DETAILS

  Remote 2014P Remote 3014P Remote 5014P Remote 3104P Remote 5104P Remote 5012PN Remote 5014PN Remote 5102PN Remote 5104PN

Specifications

Size Range 0.2 - 2.0 μm 0.3 - 25.0 μm  0.5 - 25.0 μm 0.3 - 25.0 μm 0.5 - 25.0 μm
Channel Sizes Standard: 0.2, 0.3, 0.5, 1.0 μm Standard: 0.3, 0.5, 1.0, 5.0 μm Standard: 0.5, 1.0, 5.0, 10.0 μm Standard: 0.3, 0.5, 1.0, 5.0 μm Standard: 0.5, 1.0, 5.0, 10.0 μm Standard: 0.5, 5.0μm
Optional: 0.5, 10.0μm
Counting Efficiency 50% @ 0.2 μm; 100% for particles
> 0.3 μm (per ISO 21501-4)
50% @ 0.3 μm; 100% for particles
> 0.45 μm (per ISO 21501-4)
50% @ 0.5μm; 100% for particles > 0.75μm (per ISO 21501-4) 50% @ 0.3 μm; 100% for particles > 0.45 μm (per
JIS)
50% @ 0.5 μm; 100% for particles > 0.75 μm (per JIS) 50% @ 0.5μm, 100% for particles >0.75μm (per JIS) 50% @ 0.5μm, 100% for particles >0.75μm (per ISO 21501-4) 50% @ 0.5μm, 100% for particles >0.75μm (per JIS) 50% @ 0.5μm, 100% for particles >0.75μm (per ISO 21501-4)
Flow Rate 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM) 0.1 CFM (2.83 LPM) 1.0 CFM (28.3 LPM)
Laser Source Laser Diode
Zero Count Level <1 count / 5 minutes (per ISO 21501-4) <1 count / 5 minutes (per JIS); Zero Count with Vibration: 0.02g2/Hz from 8-66 Hz <1 count / 5 minutes (per JIS)
Concentration Limits 2,000,000 Particles / ft3 @ 5% Coincidence Loss 500,000 Particles / ft3 @ 5% Coincidence Loss 2,000,000 particles / ft³ @ 5% coincidence loss 500,000 particles / ft³ @ 5% coincidence loss
Calibration Meets ISO 21501-4 Calibration using NIST Traceable PSL Spheres, DMA and Condensation Particle Counter NIST Traceable
Sample Output - Internally filtered to HEPA standards (>99.997% @ 0.3μm)
Vacuum Requirements Internal Vacuum Pump
Air Flow Internally Monitored and Controlled
Communication Modes RS-485 MODBUS, Ethernet, USB Ethernet, RS-485 MODBUS, Serial USB 4-20mA Output MODBUS: RS-485; Optional Ethernet MODBUS TCP 4-20mA Output MODBUS: RS-485; Optional Ethernet MODBUS TCP
LED Indicators Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling Power, Service Power, Flow, Service, Sampling
Data Storage 3,000 Samples - Rotating Buffer, 2.000 Records - Rotating Buffer, 2.000 Records
Connector RJ45, USB (Type B) 2 RJ45, 2 RJ12 and one USB ‘B’ 8-pin Phoenix RJ45
External Pump Switch - 2-pin Phoenix dry-contact relay - -
Supporting Software LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and LMSNet  LMS XChange; Lighthouse Monitoring System; LMS Express, Express RT and RT+; LMSNet  Lighthouse Monitoring System; LMS Xchange Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT Lighthouse Monitoring System; LMS Xchange Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT
Up to Four 4-20 mA Inputs Can be used with the Lighthouse standard 4-20 mA Temperature/Relative Humidity, Air Velocity or Differential Pressure Probes or any other 4-20 mA device -
External Alarm Output - Normally Open Dry Contact Rated 0-60 V AC/DC 1 Amp -
Enclosure Stainless Steel NEMA Style 316L Stainless Steel or Powder Coated Steel
Power 24 VDC 110 / 220 VAC, 50-60Hz or 24VDC 100W
Dimensions 7.25”(l) x 3.75”(w) x 6”(h) [18.41 x 9.52 x 15.24 cm]  9.5”(l) x 6.5”(w) x 8.51”(h) [24.13 x 16.51 x 21.62 cm]  10” (w) x 12” (h) x 6.9” (d) [25.4 x 30.4 x 14.9 cm]
Weight 5.0 lbs (2.26 kg) 12.33 lbs (5.59 kg) AC Model, 20 lbs (9 kg); DC Model, 19.5 lbs (8.85 kg)

Environmental Conditions

Operating 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Accessories

Included Operating Manual on CD; LMS XChange; Power Supply; Power Cord     Operating Manual on CD; AC or DC Power Input Connector Kit Operating Manual on CD; AC or DC Power Input Connector Kit;
Industrial RJ45 Kit; Fuse(s) 
Optional  Printed Operating Manual; REMOTE Display; REMOTE Display 10-foot Extension Cable; Network Adapter; Wireless Network Adapter; Isokinetic Sampling Probe; Sample Tubing; Data Cabling; Purge Filter; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; RS-232 Interface Cable and Adapter; High Pressure Diffuser (except 2014P); LMS Express, Express RT, RT+ and LMSNet software Printed Operating Manual; Data Cable; Sample Tubing; 24VDC 100W Power Adapter (DC Model) 

GET A QUOTE

First Name (*)
Last Name (*)
Company Name (*)
Email (*)
Phone (*)
Address (*)
Message (*)
Security Code
Security Code ImageRefresh Button
(*)
(*) Required Fields
Remote 3104P Image

Incorporating the latest in laser optical particle counting technology, Lighthouse designed the Remote 3104P airborne particle counter with pump, for continuous trouble free operation.

With a minimum sensitivity of 0.3 µm and a flow rate of 1.0 CFM (28.3 LPM), the Remote 3104P provides real-time continuous data collection at a cost-effective price per point. Using an internal vacuum source, the compact Remote 3104P particle counter can be installed where space is at a premium.

The Remote 3104P will integrate seamlessly into large facility monitoring and management systems and transfer up to 6 channels of simultaneous particle count data using RS-485 Modbus.

 

Key Features:

  • 0.3 - 25μm Range
  • 1.0 CFM (28.3 LPM) Flow Rate with Internal Pump
  • Up to 6 Channels of Simultaneous Data
  • Meets JIS Standards and ISO 21501-4 Standards
  • Compact Size
  • Stainless Steel Enclosure
  • Up to Four 4-20 mA Inputs (30 mW)
  • LED Status Indicators
  • External Alarm Connector
  • RS-485 Modbus
  • Facility Monitoring System Interface
  • Flow-controlled Internal Vacuum Pump
  • Designed for Reliability