Compatible with VHP, the REMOTE 5104V meets the stringent requirements of today’s aseptic processing applications. In addition, the REMOTE 5104V was designed to operate in CLASS 1 DIV 2 hazardous environments found in many challenging applications.
The REMOTE 5104V uses a critical orifice and external vacuum source for flow control. Built-in flow monitoring assures accuracy and reliability of the sensor’s performance.
With a sensitivity of 0.5 micron and a monitored flow rate of 1.0 CFM (28.3 LPM), the REMOTE 5104V provides real-time continuous data collection at a cost-effective price per point.
The REMOTE 5104V integrates seamlessly with large facility monitoring and building management systems. It transfers up to 4 channels of simultaneous particle count data using RS-485 Modbus.
Remote 5104V Datasheets
Letter Size – Lo Res A4 Size – Lo Res
Letter Size – Hi Res A4 Size – Hi Res
Designed and built by Lighthouse Worldwide Solutions, a name you can trust.
- VHP Compatible
- Class 1 Division 2 Rating
- Built-in Flow Monitoring
- 0.5 - 5.0 μm (optional channels up to 25 μm)
- 1.0 CFM (28.3 LPM) Flow Rate
- Up to 4 Channels of Simultaneous Data
- Meets or Exceeds JIS Standards
- Real-time Continuous Monitoring
- Small Size
- 316L Stainless Steel, Wash Down Resistant/Splash-Tight Enclosure
- Status Indicators
- RS-485 Modbus
- Interfaces to Facility Monitoring Systems
- Uses External Vacuum Source
- Designed for Reliability
- 2 Year Warranty
- International Support
- Versatile Mounting Options
- Easy to Install when Space is an Issue
- Cost Effective Price per Point
- Proven Data Integrity
- Simple Integration into FMS Systems
- Low Cost of Ownership
- Glass Coating
- Aseptic Processing
- Isolators
- Cleanrooms
Industries
- Aerospace Industry
- Pharmaceutical Industry
- Biotechnology
- Medical Applications
Specifications
| Size Range: |
0.5 - 25.0 µm |
| Channel Sizes: |
Standard: 0.5, 5.0 µm Optional 4 Channels: 0.5, 1.0, 5.0, 25.0 µm |
| Counting Efficiency: |
50% @ 0.5μm; 100% for particles > 0.75μm (per JIS) |
| Flow Rate: |
1.0 CFM (28.3 LPM) |
| Laser Source: |
Laser Diode |
| Zero Count Level: |
|
| Concentration Limits: |
500,000 Particles/ft3 @ 5% Coincidence Loss |
| Calibration: |
NIST Traceable |
| Vacuum Requirements: |
External Vacuum >18" (45.72) of Hg |
| Air Flow: |
Internally Monitored |
| Communication Modes: |
RS-485/MODBUS |
| LED Indicators: |
Power, Flow, Service, Sampling |
| Connector: |
Industrial RJ45 |
| Supporting Software: |
Lighthouse Monitoring System, LMS Express, LMS Express RT, LMS Express RT Plus, LMS Net |
| Enclosure: |
316L Stainless Steel, Wash Down Resistant/Splash-Tight |
| Power: |
24 VDC |
| Dimensions: |
4.2” (l) x 2.0” (w) x 3.7” (h) [10.7 x 5.08 x 9.398 cm] |
| Weight: |
15.6 oz (0.44 kg) |
Conditions
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
Accessories
| Included: |
Operating Manual on CD |
| Optional: |
Isokinetic Sampling Probe, Sample Tubing, Cabling, Purge Filter, Power Supply |
 |