Incorporating the latest in laser optical particle counting technology, Lighthouse designed the REMOTE 5104 for continuous trouble-free operation.
With a sensitivity of 0.5, at a flow rate of 1.0 CFM (28.3 LPM), the REMOTE 5104 provide real-time continuous data collection at a cost-effective price per point.
Using an external vacuum source, the compact REMOTE 5104 provides versatile mounting options and can be installed where space is at a premium. The optional T/RH Probe can be mounted up to eight feet from the instrument by using an optional extension cable.
The REMOTE 5104 integrates seamlessly into large facility monitoring and management systems and transfer up to 4 channels of simultaneous particle count and environmental data using RS-485 Modbus.
With the optional Network or Wireless Adapter, the REMOTE 5104 can be connected to TCP/IP LANs.
Designed and built by Lighthouse, a name you can trust.
 |
 |
 |
 |
 |
 |
| Optional T/RH Probe |
Direct Mount Isokinetic Probe |
RJ45 Connector |
LED Status |
Alarm Relay |
Flow Status Monitoring |
Remote 5104 Datasheets
Letter Size – Hi Res A4 Size – Hi Res
- 0.5 - 25μm Sensitivity
- 1.0 CFM (28.3 LPM) Flow Rate
- Up to 4 Channels of Simultaneous Data
- All Models Meet ISO 21501-4 Standards
- Compact Size
- Stainless Steel Enclosure
- Status Indicators
- External Alarm Connector
- Optional Temperature/Relative Humidity Probe
- RS-485 Modbus Output
- Facility Monitoring Systems Interface
- Versatile Mounting Options
- External Vacuum Source
- Designed for Reliability
- 2 Year Warranty
- Installs in Limited Space
- Cost Effective Price per Point
- Proven Data Integrity
- Proven Reliability
- Continuous Real-Time Data
- Easy System Integration
- International Support
- Low Cost of Ownership
Specifications
| Size Range: |
0.5 - 25μm |
| Channel Sizes: |
Standard: 0.5, 5.0μm; 0.5, 10.0μm Optional: 0.5, 1.0, 5.0, 10.0μm; 0.5, 5.0, 10.0, 25.0μm |
| Counting Efficiency: |
50% @ 0.5 μm; 100% for particles > 0.75 μm (per ISO 21501-4) |
| Flow Rate: |
1.0 CFM (28.3 LPM) |
| Laser Source: |
Laser Diode |
| Zero Count Level: |
< 1 count / 5 minutes (per ISO 21501-4) |
| Concentration Limits: |
500,000 Particles / ft3 @ 5% Coincidence Loss |
| Calibration: |
NIST Traceable |
| Data Storage: |
Rotating Buffer, 2000 Records |
| Vacuum Requirements: |
External Vacuum >18” (45.72) of Hg |
| Air Flow: |
Internally Monitored |
| Communication Modes: |
RS-485 MODBUS |
| LED Indicators: |
Power, Flow, Service, Sampling |
| Connector: |
RJ45 |
| Supporting Software: |
LMS XChange, Lighthouse Monitoring System, LMS Express, LMS Express RT, LMS Net |
| Environmental Sensors: |
(Optional) Temperature/Relative Humidity Probe 32 - 122F (0-50C)+/-1F (0.5C), 0 - 100% RH +/-2% |
| External Alarm Output: |
Normally Open Dry Contact Rated 0-60 V AC/DC 1 Amp |
| Enclosure: |
Stainless steel |
| Power: |
24 VDC |
| Dimensions: |
1.7"(l) x 4.2"(w) x 3"(h) [4.3 x 10.7 x 7.6 cm] |
| Weight: |
0.97 lb (0.44 kg) |
Conditions
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
Accessories
| Included: |
Operating Manual on CD |
| Optional: |
Printed Operating Manual; RS-485 Hub with Power Supply; RS-232 Communication Cable and Adapter; USB to Serial Port Converter; Network Adapter; Wireless Network Adapter; Isokinetic Sampling Probes; Vacuum Tubing; Sample Tubing; Data Cable; Purge Filter; Power Supply; Temperature/ Relative Humidity Probe; 8’ T/RH Probe Extension Cable; REMOTE Mounting Kit; LMS XChange, LMS Express; Express RT, LMSNet software; RS-485 Communication Kit; High-Pressure Diffuser. |