MODBUS Output
The REMOTE 5014PN meets the stringent requirements of today’s aseptic processing applications. Incorporating the latest in laser/optical particle counting technology, Lighthouse designed the REMOTE 5014PN for continuous trouble-free operation.
The REMOTE 5014PN uses a built-in vacuum pump and flow control system to insure accurate and reliable sensor performance.
With a sensitivity of 0.5 micron at a controlled flow rate of 0.1 CFM, the REMOTE 5014PN provides continuous realtime data at a cost-effective price per point.
The REMOTE 5014PN integrates seamlessly into large facility monitoring and management systems and transfers up to 4 channels of simultaneous particle count data using RS-485 MODBUS or Ethernet MODBUS TCP.
Designed and built by Lighthouse, a name you can trust.
REMOTE 5014PN Datasheets
Letter Size – Lo Res A4 Size – Lo Res
- Channel sizes: 0.5 - 5.0μm for GMP Compliance
- 0.1 CFM (2.83 LPM) Flow Rate
- Internal Vacuum Pump with Flow Control
- Up to 2 Channels of Simultaneous Data
- Meets JIS Standards
- NEMA Style 316L Stainless Steel Enclosure or
- NEMA Style Powder Coated Steel Enclosure
- LED Status Indicators
- Interfaces to Facility Monitoring Systems
- RS-485 MODBUS or Optional Ethernet MODBUS TCP Output
- 120VAC or 24VDC, 100W Models Available
- 2 Year Warranty
- Cost Effective Price per Point
- Easy System Integration
- AC and DC Power Options
- International Sales and Service Support
- Low Cost of Ownership
- HVAC Monitoring
- Indoor Air Quality Monitoring
- Food Process Monitoring
- Public and Private Building Monitoring
- Aseptic Process Monitoring
- Isolator Monitoring
- Aerospace
- Chemical Manufacturing
- Equipment Monitoring
- Pharmaceutical Packaging and Manufacturing
- Trend Analysis
- Cleanroom Laundries
- Cleanroom Packaging
- Semiconductor
- Hard Disk Drive Manufacturing
Specifications
| Size Range: |
0.5 - 25.0μm |
| Channel Sizes: |
Standard: 0.5, 5.0μm Optional: 0.5, 10.0μm |
| Counting Efficiency: |
50% @ 0.5μm, 100% for particles >0.75μm (per JIS) |
| Flow Rate: |
0.1 CFM (2.83 LPM) |
| Laser Source: |
Laser Diode |
| Zero Count Level: |
|
| Concentration Limits: |
2,000,000 particles / ft³ @ 5% coincidence loss |
| Calibration: |
NIST Traceable |
| Sample Output: |
Internally filtered to HEPA standards (>99.997% @ 0.3μm) |
| Vacuum Source: |
Internal Pump |
| Air Flow: |
Internally Monitored and Controlled |
| Communication Modes: |
MODBUS: RS-485; Optional Ethernet MODBUS TCP |
| LED Indicators: |
Power, Flow, Service, Sampling |
| I/O Connection: |
RJ-45 |
| External Pump Switch: |
2-pin Phoenix dry-contact relay |
| Supporting Software: |
Lighthouse Monitoring System; LMS XChange; LMS Express, Express RT |
| Enclosure: |
NEMA Style 316L Stainless Steel or Powder Coated Steel |
| Power: |
110 / 220 VAC, 50-60Hz or 24VDC 100W |
| Dimensions: |
10” (w) x 12” (h) x 6.9” (d) [25.4 x 30.4 x 14.9 cm] |
| Weight: |
AC Model, 20 lbs (9 kg); DC Model, 19.5lbs (8.85 kg) |
Conditions
| Operating: |
50°F to 104°F (10°C to 40°C), Humidity 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C), Humidity up to 98% non-condensing |
Accessories
| Included: |
Operating Manual on CD; AC or DC Power Input Connector Kit; Industrial RJ45 Kit; Fuse(s) |
| Optional: |
Printed Operating Manual; Cabling; Data Cable; Sample Tubing; 24VDC 100W Power Adapter (DC Model) |