4-20mA Output
The REMOTE 5012PN meets the stringent requirements of today’s aseptic processing applications. Incorporating the latest in laser optical particle counting technology, Lighthouse designed the REMOTE 5012PN for continuous trouble-free operation.
The REMOTE 5012PN uses a built-in vacuum pump and flow control system to insure accurate and reliable sensor performance.
With a sensitivity of 0.5 micron at a controlled flow rate of 0.1 CFM, the REMOTE 5012PN provides continuous realtime data at a cost-effective price per point.
The REMOTE 5012PN integrates seamlessly into large facility monitoring and management systems and transfers up to 2 channels of simultaneous particle count data using 4-20mA output.
Designed and built by Lighthouse, a name you can trust.
REMOTE 5012PN Datasheet
Letter Size – Lo Res A4 Size – Lo Res
- Channel sizes: 0.5 - 5.0μm for GMP Compliance
- 0.1 CFM (2.83 LPM) Flow Rate
- Internal Vacuum Pump with Flow Control
- Up to 2 Channels of Simultaneous Data
- Meets JIS Standards
- NEMA Style 316L Stainless Steel Enclosure or
- NEMA Style Powder-coated Steel Enclosure
- LED Status Indicators
- Interfaces to Facility Monitoring Systems
- 4-20mA Output
- 120VAC or 24VDC, 100W Models Available
- 2 Year Warranty
- Cost Effective Price per Point
- Easy System Integration
- AC and DC Power Options
- International Sales and Service Support
- Low Cost of Ownership
- HVAC Monitoring
- Indoor Air Quality Monitoring
- Food Process Monitoring
- Public and Private Building Monitoring
- Aseptic Process Monitoring
- Isolator Monitoring
- Aerospace
- Chemical Manufacturing
- Equipment Monitoring
- Pharmaceutical Packaging and Manufacturing
- Trend Analysis
- Cleanroom Laundries
- Cleanroom Packaging
- Semiconductor
- Hard Disk Drive Manufacturing
Specifications
| Size Range: |
0.5 - 25.0μm |
| Channel Sizes: |
Standard: 0.5, 5.0μm Optional: 0.5, 10.0μm |
| Counting Efficiency: |
50% @ 0.5μm, 100% for particles >0.75μm (per JIS) |
| Flow Rate: |
0.1 CFM (2.83 LPM) |
| Laser Source: |
Laser Diode |
| Zero Count Level: |
|
| Concentration Limits: |
2,000,000 particles / ft³ @ 5% coincidence loss |
| Calibration: |
NIST Traceable |
| Sample Output: |
Internally filtered to HEPA standards (>99.997% @ 0.3μm) |
| Vacuum Source: |
Internal Pump |
| Air Flow: |
Internally Monitored and Controlled |
| Communication Modes: |
4-20mA Output |
| LED Indicators: |
Power, Service |
| I/O Connection: |
8-pin Phoenix |
| External Pump Switch: |
2-pin Phoenix dry-contact relay |
| Supporting Software: |
Lighthouse Monitoring System; LMS XChange |
| Enclosure: |
NEMA Style 316L Stainless Steel or Powder Coated Steel |
| Power: |
110 / 220 VAC, 50-60Hz or 24VDC 100W |
| Dimensions: |
10” (w) x 12” (h) x 6.9” (d) [25.4 x 30.4 x 14.9 cm] |
| Weight: |
AC Model, 20 lbs (9 kg); DC Model, 19.5 lbs (8.85 kg) |
Conditions
| Operating: |
50°F to 104°F (10°C to 40°C), Humidity 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C), Humidity up to 98% non-condensing |
Accessories
| Included: |
Operating Manual on CD; AC or DC Power Input Connector Kit |
| Optional: |
Printed Operating Manual; Data Cable; Sample Tubing; 24VDC 100W Power Adapter (DC Model) |