Using state-of-the-art laser technology, the Lighthouse REMOTE 1104LD is an industry first – a laser diode based remote particle counter with a sensitivity of 0.10 micron and a high flow rate of 1.0 CFM (28.3 LPM).
Designed with reliability and quality in mind, the REMOTE 1104LD features our long-life laser diode technology that provides higher concentration limits, a better signal to noise ratio and requires less routine maintenance than conventional HeNe laser-based particle counters.
The REMOTE 1104LD supplies continuous real time data at a cost-effective price per point. Very simple to integrate with large facility monitoring and management systems, the REMOTE 1104 can transfer up to 8 channels of simultaneous particle count data via RS-485 MODBUS.
Designed and built by Lighthouse, a name you can trust.
Remote 1104LD Datasheets
Letter Size – Lo Res A4 Size – Lo Res
- 0.10 - 1.0 μm Sensitivity
- 1.0 CFM (28.3 LPM) Flow Rate
- Up to 8 Channels of Simultaneous Data
- World’s Smallest 1.0 CFM 0.10 µm counter
- Meets ISO 21501-4 Standards
- Long Life Laser Diode
- Small Footprint
- Status Indicators
- RS-485 MODBUS
- Stainless Steel Enclosure
- Interfaces to Facility Monitoring Systems
- Uses External Vacuum Source
- Designed for Reliability
- Designed to Ensure 21 CFR Part 11 Compliance when used with Additional Software
- No Routine Laser Tube Replacements
- No Routine Laser Optics Cleaning
- 2 Year Warranty
- Cost Effective Price per Point
- Proven Data Integrity
- Provides Continuous Real-Time Data
- Reduced Maintenance costs over HeNe
- Easy Systems Integration
- International Support
- Low Cost of Ownership
Specifications
| Size Range: |
0.10 - 1.0 μm |
| Channel Sizes: |
Standard: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0 μm |
| Communication Modes: |
RS 485 MODBUS |
| Counting Efficiency: |
50% @ 0.1 μm; 100% for particles > 0.15 μm (per JIS) |
| Flow Rate: |
1.0 CFM (28.3 LPM) |
| Laser Source: |
Long Life Laser Diode |
| Zero Count Level: |
|
| Concentration Limits: |
500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss, per ISO 21501-4 |
| Calibration: |
NIST Traceable, per ISO 21501-4 |
| Vacuum Requirements: |
External Vacuum required, nominal 1.0 CFM >18" (45.72 cm) of Hg |
| Led Indicators: |
Power, Flow, Service, Sampling |
| Supporting Software: |
Lighthouse Monitoring System, LMS Express, LMS Express RT, LMS Express RT Plus |
| Enclosure: |
Stainless steel |
| Power: |
24 VDC |
| Dimensions: |
10.5” (w) x 7.65” (h) x 3.35” (d) [26.67 x 19.43 x 8.51 cm] |
| Weight: |
7.0 lb (3.18 kg) |
Conditions
| Operating: |
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing |
| Storage: |
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing |
Accessories
| Included: |
Operating Manual on CD, Purge Filter, Power Supply, Power Cord |
| Optional: |
Isokinetic Sampling Probe, Sample Tubing, Cabling |
 |